ATP-parameter
ATP-One-Shot-parameter; ATP i VF
Progressivt behandlingsforløp
Postsjokk-stimulering
Parameter
Område
Standard
ATP-type
Burst; Ramp; Burst + PES
Burst
ATP-forsøk
OFF
1...(1)...10
3
Ventrikkelstimulering
(kun HF-implantater)
RV; LV; BiV
RV
S1-antall
1...(1)...10
5
Ekstra S1
OFF; ON
ON
R-S1-intervall
200...(10)...500 ms (absolutt)
70...(5)...95 % (adaptiv)
80 %
S1-dekrement
5...(5)...40 ms
10 ms
Scan-dekrement
OFF
5...(5)...40 ms
OFF
S1-S2-intervall
200...(10)...500 ms (absolutt);
70...(5)...95 % (adaptiv)
70 %
Minimalt ATP-intervall
200...(5)...300 ms
200 ms
ATP-timeout OFF
00:15... (00:15) ...05:00 mm:ss
OFF
ATP-optimering
OFF; ON
OFF
ATP-impulsamplitude
7,5 V
7,5 V
ATP-impulsvarighet
1,5 ms
1,5 ms
Parameter
Område
Standard
ATP-type
OFF;
Burst; Ramp; Burst + PES
OFF
ATP-forsøk
1
1
Ventrikkelstimulering
(kun HF-implantater)
RV; LV; BiV
RV
S1-antall
1...(1)...10
8
R-S1-intervall
200...(10)...350 ms (absolutt)
70...(5)...95 % (adaptiv)
85 %
S1-dekrement
5...(5)...40 ms
10 ms
S1-S2-intervall
200...(10)...350 ms (absolutt)
70...(5)...95 % (adaptiv)
70 %
Stabilitet
12 %
12 %
ATP-impulsamplitude
7,5 V
7,5 V
ATP-impulsvarighet
1,5 ms
1,5 ms
Parameter
Område
Standard
Progressivt behandlingsforløp
OFF; ON
ON
Parameter
Område
Standard
Modus (automatisk
programmert)
• DDI ved permanent DDD(R),
DDI(R), AAI(R)
• VDI ved permanent VDD(R),
VDI(R)
• VVI ved permanent VVI(R);
• OFF
Grunnfrekvens
30...(5)...100... (10)...160 ppm
60 ppm
Frekvenshysterese
OFF
-5...(-5)... -65 ppm
OFF
AV-tid
50...(10)...100... (10)...160 ms
140 ms
Ventrikkelstimulering
RV; BiV
RV
LV T-bølgebeskyttelse
OFF; ON
ON
Trigging
OFF
RVs, RVs + VES
OFF
Postsjokk-varighet
OFF
00:10...(00:10)... 00:50;
01:00...(01:00)... 10:00 mm:ss
00:10 mm:ss
Parameter
Område
Standard