Forced termination; with SMART, SMART redetection
Redetection counter
Therapy Parameters
Ventricle
ATP parameters
ATP One Shot parameters; ATP in VF
Parameter
Range
Standard
Forced termination
OFF 1 ... (1) ... 15 min
1 min
Parameter
Range
Standard
Redetection counter VT1
10 ... (2) ... 30
20
Redetection counter VT2
10 ... (2) ... 30
14
Parameter
Range
Standard
Energy of 1st shock and 2nd
shock VT1, VT2
OFF 1 ... (1) ... 16 ... (2) ... 40 J
40 J
Energy of 1st shock and 2nd
shock VF
1 ... (1) ... 16 ... (2) ... 40 J
40 J
Shock sequences 3rd-nth shock
in VT1
OFF
1*40 J; 2*40 J; 3*40 J; 4*40 J;
5*40 J; 6*40 J
6*40 J
Shock sequences 3rd-nth shock
in VF
OFF
4*40 J; 5*40 J; 6*40 J
6*40 J
Number of shocks
(VT1/VT2)
0 ... (1) ... 8
8
Number of shocks (VF)
2; 6 ... (1) ... 8
8
Confirmation in VT1, VT2, VF
OFF; ON
ON
Shock form in VT1, VT2, VF
Biphasic; biphasic2
Biphasic
Polarity in VT1, VT2, VF
Normal; inverse; alternating
Normal
Shock path (valid for all shocks
including the painless shock
impedance)
RV
SVC + housing
RV
housing
RV
SVC
RV
SVC +
housing
Shock path DX
RV
housing
Parameter
Range
Standard
ATP type
Burst; Ramp; Burst + PES
Burst
ATP attempts
OFF
1 ... (1) ... 10
3
Ventricular pacing
(HF devices only)
RV; LV; BiV
RV
Number S1
1 ... (1) ... 10
5
Add. S1
OFF; ON
ON
R-S1 interval
200 ... (10) ... 500 ms (absolute);
70 ... (5) ... 95% (adaptive)
80%
S1 decrement
5 ... (5) ... 40 ms
10 ms
Scan decrement
OFF
5 ... (5) ... 40 ms
OFF
S1-S2 interval
200 ... (10) ... 500 ms (absolut);
70 ... (5) ... 95% (adaptive)
70%
Minimum ATP interval
200 ... (5) ... 300 ms
200 ms
ATP timeout
OFF
00:15 ... (00:15) ... 05:00 mm:ss
OFF
ATP optimization
OFF; ON
OFF
ATP pulse amplitude
7.5 V
7.5 V
ATP pulse width
1.5 ms
1.5 ms
Parameter
Range
Standard
ATP type
OFF;
Burst; Ramp; Burst + PES
OFF
ATP attempts
1
1
Ventricular pacing
(HF devices only)
RV; LV; BiV
RV
Number S1
1 ... (1) ... 10
8
387158--G_GA_Lumax540_mul.book Page 19 Friday, February 1, 2013 5:38 PM