background image

14

(3) LCD display:

 A "Supertwist" liquid crystal 

display (LCD) meter allows display of readings, 
messages, and other information. 

(4) Alarm lights:

 Four LED (light emitting diode) 

alarm lights provide a visual indication of alarm 
state.  Each light is dedicated to a single 
channel of detection, and will emit a bright red 
light when a sensor alarm level is exceeded.  
One of the alarm light assemblies also includes 
a photo-sensor used to monitor the level of 
background illumination.  An automatically-
activated backlight brightens the meter display 
whenever the instrument is taken into a dark 
area. 

(5)  On / Off "mode" button:

 The large black push-

button is called the "mode" button.  It is used to 
turn the PhD Ultra on and off and to control 
most other operations, including the automatic 
calibration adjustment. 

(6)  Sensor compartment cover:

 The sensors are 

protected by a vented sensor compartment 
cover.  Its own protective filter cap individually 
protects each sensor.  A water-resistant PVC 
gasket and inner-liner protect the instrument 
against leakage or exposure to liquids.  

(7)  Audible alarm orifice:

 A cylindrical orifice 

extending through the sensor compartment 
cover houses the audible alarm.  The waterproof 
audible alarm seats directly to the PVC inner-
liner to protect the instrument against leakage or 
exposure to liquids. 

(8) Battery pack:

 Two types of interchangeable 

battery packs (rechargeable NiCad and 
disposable alkaline) are available for use in the 
PhD Ultra.  NiCad battery packs may be 
recharged while the pack is installed in the 
instrument, or the battery pack may be removed 
from the instrument and recharged separately. 

Chapter 2 of the PhD Ultra owner's manual 
covers battery replacement and charging 
procedures. 

(9)  Battery charger connector: 

A water-resistant 

connector at the rear of the case is used to 
connect the PhD Ultra to the “drop in” style PhD 
Ultra fast charger. 

(10) Bottom 

surface:

 A sturdy clip allows the 

user to wear the PhD Ultra on a belt or other 
article of clothing.  Snaps are provided to hold 
the PhD Ultra securely in the padded leather 
weather cover. 

(11) Key 

pad:

 The key pad is located on the 

back of the PhD Ultra under the belt clip.  Slide 
the belt clip towards the rear of the instrument to 
access the four small push-buttons.  The key 
pad consists of, four buttons labeled  “+”, “-”, 
“CAL”, and "ALM”.  These buttons are normally 
only used during set-up and other programming 
procedures.   

Use of these push buttons is reserved for 
authorized personnel.

1.11  PhD Ultra accessories 

Each PhD Ultra is delivered in a foam lined box 
containing the PhD Ultra detector, padded leather 
weather cover, carrying strap, sample draw / 
calibration adopter, hand-aspirated sample-draw kit 
(with 10 feet of sample draw tubing), 2 feet of 
additional tubing for use during calibration, 
reference manual, quick reference card, and 
comprehensive training video. 

The sample draw kit comprises a slip-on sample 
draw / calibration adapter, squeeze bulb, sample 
probe, replacement sample probe filters, and ten 
feet of tubing. 

1.11.1  “Alkaline” PhD Ultra detectors   

If the PhD Ultra has been purchased as an “alkaline” 
instrument the standard accessories also include an 
alkaline battery pack and a set of 3 disposable AA 
alkaline batteries. 

1.11.2  “NiCad” PhD Ultra detectors   

If the PhD Ultra has been purchased as a “NiCad” 
instrument the standard accessories additionally 
include a NiCad battery pack, slip in PhD Ultra fast 
charger, as well as a spare alkaline battery pack and 
set of disposable batteries. 

1.11  PhD Ultra kits 

PhD Ultra detectors may also be purchased as part 
of complete kits.   

1.11.1  PhD Ultra Confined Space Kits 

Besides the standard accessories included with 
every PhD Ultra, Confined Space Kits also include 
calibration fittings, regulator, one cylinder of each of 
the appropriate calibration gases and foam lined 
waterproof carrying case for the instrument, 
calibration materials and other accessories.

 

Cylinders used in Confined Space Kits contain 
either 58 or 103 liters depending on the specific 
mixture of gas. 

1.11.2  PhD Ultra Value Packs 

It is also possible to order the PhD Ultra configured 
as a “Value Pack”.  The Value Pack comprises an 
alkaline PhD Ultra, all standard accessories, 34 liter 
cylinder of all-in-one calibration gas, fixed flow rate 
regulator, and foam lined carrying case.

 

Содержание PhD Ultra

Страница 1: ...Reference Manual PhD Ultra Multi Gas Detector 651 South Main Street Middletown CT 06457 USA 860 344 1079 800 711 6776 FAX 860 344 1068 http www biosystems com Version 2 60 28JAN2008 Part Number 13 03...

Страница 2: ...NS IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED AVERTISSEMENT LIRE ATTENTIVEMENT LES INSTRUCTIONS AVANT DE METTRE EN MARCHE PhD Ultra Operation Manual Ver...

Страница 3: ...6 TEMPERATURE OUT OF RANGE 12 1 6 4 7 OTHER ELECTRONIC SAFEGUARDS 12 1 6 4 8 SECURITY BEEP 12 1 7 INSTRUMENT FIRMWARE REQUIREMENT FOR COMPATIBILITY WITH THE IQ SYSTEM 12 1 8 CLASSIFICATION FOR INTRIN...

Страница 4: ...2 4 3 2 CHANGING SAMPLE PROBE TUBES 23 2 5 BIOSYSTEMS EEPROM EQUIPPED SMART SENSORS 24 2 5 1 Identification of type of sensor by instrument 24 2 5 2 Other information stored with the sensor EEPROM 24...

Страница 5: ...38 4 4 RE INITIALIZING THE PHD ULTRA 38 4 5 RECORD KEEPING 39 4 5 1 PhD Ultra datalogging overview 39 4 5 2 Optional Datalink and Gas Detection Database Software Kit 39 4 5 3 Adjusting record keeping...

Страница 6: ...IZED PUMP MAINTENANCE 49 5 4 1 Internal pump filter replacement 50 5 4 2 Specific problems with motorized pump 51 5 4 2 1 PUMP WILL NOT TURN ON 51 5 4 2 2 CAN T RESUME NORMAL OPERATION AFTER A LOW FLO...

Страница 7: ...for combustible gas monitoring in atmospheres where oxygen concentrations exceed 21 0 oxygen ONLY THE COMBUSTIBLE GAS DETECTION PORTION OF THIS INSTRUMENT HAS BEEN ASSESSED FOR PERFORMANCE BY CSA UNIQ...

Страница 8: ...An instrument that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored and the instrument is once again fit for use 7 Do not r...

Страница 9: ...ectrochemical toxic sensors that have been designed to minimize the effects of common interfering gases These sensors provide accurate dependable readings for toxic gases commonly encountered during c...

Страница 10: ...c reading may indicate a hazardous combustible gas concentration that exceeds the PhD Ultra s zero to 100 percent LEL detection range Failure to immediately leave the area may result in serious injury...

Страница 11: ...potentially life threatening atmospheric conditions Any alarm condition should be taken seriously The safest course of action is to immediately leave the affected area and return only after further te...

Страница 12: ...nt firmware version is given immediately after the PhD Ultra is turned on 1 8 Classification for intrinsic safety The PhD Ultra is Classified by Underwriters Laboratories Inc and the Canadian Standard...

Страница 13: ...nt is being used in a hazardous location Alkaline battery packs may not be opened and alkaline batteries may not be replaced while the battery pack is located in a hazardous area Battery replacement a...

Страница 14: ...he padded leather weather cover 11 Key pad The key pad is located on the back of the PhD Ultra under the belt clip Slide the belt clip towards the rear of the instrument to access the four small push...

Страница 15: ...ill then be shown Note If the date and time are incorrect see section 4 5 3 3 for instructions The PhD Ultra automatically evaluates itself to determine its electronic fitness for use by performing an...

Страница 16: ...be put back into service or used until the accuracy of any affected sensor has been verified by exposure to the appropriate known concentration test gas The Needs Cal warning message may be acknowled...

Страница 17: ...cal reading changes to reflect the new value the LED alarm light flashes and the audible alarm sounds PhD Ultra alarms are normally self resetting When readings drop below the pre set alarm levels vis...

Страница 18: ...instrument keypad Push the and buttons at the same time to change operating modes Each time the operating mode is changed the display screen will briefly indicate the operating mode that has been sel...

Страница 19: ...rinsic safety 2 3 3 Low battery alarms The PhD Ultra is designed to automatically determine which type of battery pack has been installed and use the appropriate low battery alarms Alarms will be acti...

Страница 20: ...in and check to see that the PWR power on indicator LED on the charger cradle is lit 4 Slip the PhD Ultra into the charger cradle and check to see that the FAST fast charger indicator LED on the charg...

Страница 21: ...ample draw adapter 3 Connect the other end of the hose to the sample probe as shown in Figure 2 4 1 1 4 Attach the sample draw adapter to the PhD Ultra as shown in Figure 2 4 1 2 5 Cover the end of th...

Страница 22: ...4 2 1 Using the continuous sample draw pump 1 Turn the PhD Ultra on Figure 2 4 2 1 1 Sample draw pump and probe assembly 2 Connect the slip on battery operated pump with the hose and probe assembly as...

Страница 23: ...taminants are removed by means of a cellulose filter similar to those used in filter cigarettes The hydrophobic filter includes a 0 1 Pm Teflon barrier which blocks the flow of moisture as well as any...

Страница 24: ...henever a change is made during operation and whenever the instrument is turned off If a sensor is changed or replaced the PhD Ultra notes that a change has occurred displays a Needs Cal message the n...

Страница 25: ...to inaccurate and potentially dangerous readings and may void the standard Biosystems warranty Customers are strongly urged to use only Biosystems calibration materials when calibrating the PhD Ultra...

Страница 26: ...in serious injury or death Avertissement Toute lecture rapide et positive suivie d une baisse subite au erratique de la valeur peut indiquer une concentration de gaz hors gamme de d tection qui peut...

Страница 27: ...instrument will display H2S and hydrogen sulfide must be used to verify accuracy Use of the incorrect calibration gas may lead to inaccurate and potentially dangerous readings If the sensor is calibr...

Страница 28: ...d value during a bump test fails to fall between 100 and 120 of the expected value for the gas Always check the expiration date on calibration gas cylinder s prior to use Expired calibration gas can l...

Страница 29: ...Calibration Adjustment screen This screen also includes a timer that counts down the number of seconds left to initiate the adjustment Note If the MODE button is pressed at any time prior to completi...

Страница 30: ...en the instrument has detected that the proper gas has been applied the sensor will be adjusted 6 When all sensors currently installed have been successfully span adjusted the display will announce Au...

Страница 31: ...ns Span calibration procedures using buttons on the instrument keypad are most useful when calibration of only a single sensor is desired Span calibration using the keypad buttons may only be done whi...

Страница 32: ...gas is applied to the sensor being adjusted a screen will be displayed indicating that the span gas concentration is too low Verify that the flow rate of the regulator is 1 0 liters per minute Replac...

Страница 33: ...tings can be restored at any time during normal operation by using the procedures discussed in Section 4 2 2 In many cases it is possible to comply with OSHA guidelines while using higher alarm points...

Страница 34: ...reens would be shown Oxygen and combustible gas LEL The ceiling alarm set points for the toxic sensors currently installed The STEL and TWA alarm set points for the toxic sensor currently installed in...

Страница 35: ...L button to advance the display to the next available setup option or press Mode to exit from the instrument setup screens 4 3 2 Changing the precision of the toxic sensor read out The toxic decimal p...

Страница 36: ...age showing the minutes and seconds between beeps Use the and keys to set the interval Setting the interval to 0 will turn the security beep off 4 3 7 Low temperature alarms To access the low temperat...

Страница 37: ...on Gas screen If an LEL combustible gas sensor has been installed the LEL sensor s calibration gas concentrations will be the first shown This screen indicates the concentration and type of test gas t...

Страница 38: ...3 seconds A screen will briefly announce that changes have been made and the instrument setup has been updated The PhD Ultra will then turn itself off Pressing mode button to cancel the changes and s...

Страница 39: ...ing data may be stored Datalogging is a transparent function that is continually in operation As long as the datalogger has not been disabled the instrument will begin recording data as soon as the in...

Страница 40: ...o not effect the way in which TWA STEL Ceiling and Peak exposure values are calculated To modify the datalogger interval first enter the datalogging adjust mode as described in section 4 5 3 1 Once th...

Страница 41: ...he software is set to the other proper communication will not be possible The software communication rate setting can be modified through the PC Setup screen as discussed in the PhD Ultra Datalink man...

Страница 42: ...sion Press to view the oldest monitoring session in the datalogger memory After a few seconds the PhD Ultra will automatically show the most recent monitoring session The numbers in this screen signif...

Страница 43: ...user ID is set press the mode button to display the list of locations currently in the instrument memory The first screen will show the current monitoring location on the top row with the name of the...

Страница 44: ...be displayed whenever the instrument is first turned on Pressing the mode button acknowledges the message and allows normal operation The message will continue to be displayed each time the instrumen...

Страница 45: ...e current status of the passcode option To disable the passcode option press the button To enable the passcode option press the button Press the CAL button to save the new setting Press the mode butto...

Страница 46: ...lunt tool B4 Remove sensor blank from the sensor compartment Then proceed to step C5 or D5 depending on the sensor type For Sensors O2 LEL CO CO Plus H2S NO perform step C5 C5 Place the new filter sna...

Страница 47: ...nt is connected to a 110 VAC power source through the battery charger The PhD Ultra must be located in a non hazardous location during the charging cycle Charging the PhD Ultra in a hazardous location...

Страница 48: ...Technician operating mode Take the instrument to a non hazardous location If equipped with an alkaline battery pack replace the batteries and attempt to turn the instrument back on If equipped with a...

Страница 49: ...en supplied with your replacement PROM Do not use small screwdrivers or other non recommended devices to remove the old PROM Use of non recommended devices can easily cause damage to the PROM socket C...

Страница 50: ...sections of the pump case as shown in Figure 5 4 3 2 Remove the two pieces of tubing from the filter cap by gently twisting and pulling until the ends are free 3 Remove the filter cap and replace the...

Страница 51: ...5 5 Returning your PhD Ultra to Biosystems for service or repair Please contact the Biosystems Service Department at 860 344 1079 to obtain a Return Authorization number prior to shipment A Biosystems...

Страница 52: ...rt oxygen Although very high concentrations of carbon monoxide may be acutely toxic and lead to immediate respiratory arrest or death it is the long term physiological effects due to chronic exposure...

Страница 53: ...to be present in LEL concentrations in order for this reaction to occur Even trace amounts of combustible gas present in the air surrounding the sensor will be catalytically burned on the surface of t...

Страница 54: ...1910 146 OSHA has determined that a combustible hazard exists whenever the concentration of combustible gas or vapor exceeds 10 percent LEL Per this standard confined spaces with concentrations which...

Страница 55: ...inst this potentially dangerous sort of calibration error Biosystems Propane Equivalent calibration mixtures are based on methane but in concentrations which are designed to produce a level of sensiti...

Страница 56: ...ppm 1 0 ppm HCN 4 7 ppm 4 7 ppm 4 7 ppm NH3 25 ppm 25 ppm 35 ppm NO 25 ppm 25 ppm 25 ppm NO2 1 0 ppm 1 0 ppm 1 0 ppm Oxygen O2 Low alarm 19 5 High alarm 22 0 Combustible gas 10 LEL Biosystems PhD Ult...

Страница 57: ...EZ Cal Value Pack multi component calibration gas CO 50 ppm propane equivalent 50 LEL in air Only for use with 54 12 022 Value Pack style fixed flow rate regulator 54 9041 EZ Cal multi component gas...

Страница 58: ...s has with the sensor the effect can either decrease the signal negative cross sensitivity or increase the signal positive cross sensitivity Each figure represents the reaction of the sensor to 100 pp...

Страница 59: ...closely follow these guidelines If your operating procedures do not permit daily checking of the sensors Biosystems recommends the following procedure to establish a safe and prudent accuracy check s...

Страница 60: ...ears from date of purchase Toxi Limited 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Toxi3Ltd 3 years after activation or 3 years after the Must Be Act...

Отзывы: