The Base unit has an aerator on the lower half. In addition up to 6 optional
aeration units can be used, two of which can be transported in the ProteQ. The
building’s HVAC system can also be used to aerate. The aerators contain a
catalyst, which converts the hydrogen peroxide (H
2
O
2
) vapour to water vapour
and oxygen. This phase continues until the hydrogen peroxide concentration
reaches 1ppm, when it should be checked with the hand held low level sensor.
Note:
It is against US federal law to use a biocide in a manner inconsistent
with its labelling. Refer to and comply with Bioquell Hydrogen Peroxide
sterilant labelling.