BIOQUELL PROTEQ USER MANUAL -EN
TD092-O&M-001
REVISION 5 DRAFT
Page 33 of 58
12
PACKING-UP
The unit should be packed away neatly to avoid damage.
WARNING
–
Do not enter an enclosure until the hydrogen peroxide
level has been checked with a low level sensor and is at a safe level
1.
Turn the power off to the Base unit and aeration units if
used. Coil-up the power cables and place in the storage
area by the power switch of the Base unit.
2.
Return the aeration units, if present, onto the back of
the Base unit.
3.
Wheel the Base unit out of the room to the lectern’s
location.
4.
Place the hand held low level sensor on the lectern, or
return it to its charging unit.
5.
Unplug the lectern’s power cable. DO NOT TAKE THE
LECTERN INTO THE ROOM, this may re-introduce
contamination.
6.
Place the power cable’s plug behind the grey metal as
shown and then wind the cable around the groove
being held by the grey metal.
7.
Place the lectern onto the Base unit by leaning it
backwards and first engaging the bottom and then
pushing the top forward.
The bottles do not need to be removed from the ProteQ as
long as they remain engaged with the plunger down. If
removing bottles, always fit BOTH yellow and screw caps.
Содержание ProteQ
Страница 53: ...BIOQUELL PROTEQ USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Page 53 of 58 20 SCREEN NAVIGATION...
Страница 54: ...BIOQUELL PROTEQ USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Page 54 of 58...
Страница 55: ...BIOQUELL PROTEQ USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Page 55 of 58 21 DECLARATION OF CONFORMITY...
Страница 57: ...BIOQUELL PROTEQ USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Page 57 of 58 THIS PAGE IS INTENTIALLY BLANK...