BIOQUELL PROTEQ USER MANUAL -EN
TD092-O&M-001
REVISION 5 DRAFT
Page 18 of 58
Bottles may be left in the vaporizer between decontamination cycles. Discard part
used bottles if left for more than a month.
STEP 3:
Carefully remove the internal rubber yellow cap
by pulling it off. Retain the cap.
WARNING:
Use the finger grips in the 2 litre bottle to hold the bottles
securely and safely. If using a 950ml bottle the insert must be fitted.
STEP 4:
For 2 litre bottle:
Pull the bottle tray open with one hand while
holding the bottle with the other using the
finger grips. These should be facing forward for
correct alignment. Then slide the bottle in.
For 950ml bottles:
Pull the bottle tray open with one hand while
placing the bottle in the insert in the bottle
carrier with the other. An insert must be
present, see section 7.3
Ensure that the stud at the bottom of the holder
lines-up with the indent on the bottom of the
bottle. It may be necessary to rotate the bottle
until it drops down into its correct position.
The label in the bottom of the bottle holder also
illustrates this.
STEP 5:
Push the black plunger down until it clicks and
stays down.
Note – avoid pressing the silver button during
this step, as the silver button releases the
mechanism.
Содержание ProteQ
Страница 53: ...BIOQUELL PROTEQ USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Page 53 of 58 20 SCREEN NAVIGATION...
Страница 54: ...BIOQUELL PROTEQ USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Page 54 of 58...
Страница 55: ...BIOQUELL PROTEQ USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Page 55 of 58 21 DECLARATION OF CONFORMITY...
Страница 57: ...BIOQUELL PROTEQ USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Page 57 of 58 THIS PAGE IS INTENTIALLY BLANK...