background image

 

 

BIOQUELL PROTEQ USER MANUAL -EN 

TD092-O&M-001 

REVISION 5 DRAFT 

 

 
 

Page 17 of 58 

 

 

 

 

 

 

If a Flexi Adaptor is fitted the left hand light will illuminate green and the right 

hand will be unlit. No other colours will be seen. 

 

7.2

 

Installing Hydrogen Peroxide Bottles 

 

The following instructions explaining how to install bottles must be followed; in 

addition, there is an illustration of the procedure on the vaporizer module above 

the bottle enclosure. Before loading a 950ml bottle ensure the insert is present, 

see section 7.3. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WARNING: 

Use Personal Protective Equipment when handling hydrogen 

peroxide bottles. As a minimum, wear suitable gloves and goggles.  

 

 

 

Press the bottom right hand side of the flap 

under the bottle storage compartment.  

 

The flap will spring-up. 

 

 

 

Remove a pair of gloves and goggles. Put them 

on before handling the bottle of hydrogen 

peroxide. 

 

 

Loading new bottles: 

 

STEP 1: 

To access the bottles push the silver button at 

the centre of the black plunger. The plunger will 

be released and will move up. 

STEP 2: 

Obtain a new bottle and check the outer bag for 

any signs of liquid. If liquid is present, do not 

use the bottle. Treat the liquid as hydrogen 

peroxide and take the necessary precautions. If 

no liquid is present, remove the bottle from the 

bag, and remove the lid carefully by turning 

counter clockwise. Retain the lid. 

Содержание ProteQ

Страница 1: ...Q USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Bioquell UK Ltd 52 Royce Close West Portway Andover Hampshire SP10 3TS T F E W 44 0 1264 835 835 44 0 1264 835 836 info bioquell com www bioquell com Hy...

Страница 2: ...or approved by FDA but has been authorized for emergency use by FDA under an EUA for the decontamination of compatible N95 respirators for single user reuse by HCP to prevent exposure to SARS CoV 2 a...

Страница 3: ...cle 20 8 2 Viewing Cycle Variables Sensor units only 23 8 3 Aeration Only Cycle 24 8 4 Printed Report 24 8 5 Log OFF 25 8 6 Process Stop 25 9 LOADING AN EXISTING PRE PROGRAMMED CYCLE 26 10 CREATING A...

Страница 4: ...tory symbol is used to describe other safety precautions that must be followed before operating the equipment HANDLING HYDROGEN PEROXIDE Always wear gloves long sleeves and eye protection when handlin...

Страница 5: ...contaminated A suitable portable H2O2 detector should be used throughout the cycle to monitor concentrations outside the area If leak is detected leak sealing and evacuation procedures should be follo...

Страница 6: ...ents to the Bioquell ProteQ generator must only be made by trained Bioquell personnel or their agents The equipment must be serviced in accordance with the Maintenance Schedule to maintain safe perfor...

Страница 7: ...B area WEAR PROTECTIVE GLOVES Wear protective clothing when handling hydrogen peroxide bottles WEAR EYE PROTECTION Wear protective eye protection when handling hydrogen peroxide bottles READ MANUAL Re...

Страница 8: ...he small boxes containing the hand held instruments 3 Remove the large outer sleeve by lifting it up and over the unit 4 Lower the ramp 5 Carefully cut open the silver bag 6 Carefully with two people...

Страница 9: ...itioning the dosing pump is primed with hydrogen peroxide liquid 3 1 2 Gassing Phase Liquid from a bottle is pumped onto the vaporizer heaters Air then passes around the heaters where it picks up the...

Страница 10: ...ystem can also be used to aerate The aerators contain a catalyst which converts the hydrogen peroxide H2O2 vapour to water vapour and oxygen This phase continues until the hydrogen peroxide concentrat...

Страница 11: ...eration Unit Braked Wheels 2 Top air inlet filter Air outlets 4 Bottle engagement plunger Wireless status indicator Bottle status indication lights Bottle compartment Safety equipment storage Gloves g...

Страница 12: ...uately rated cables 4 1 3 Aeration Unit Optional Air inlet filter Distribution fans Wireless status indicator Handles Power input and main switch Air outlet grille Colour touch screen Printer Wireless...

Страница 13: ...ith approved Hydrogen Peroxide Ensure eye protection and gloves are worn as a minimum and filling takes place in a suitably ventilated area 5 MOVING THE PROTEQ WARNING Personal Protective Equipment mu...

Страница 14: ...ectern and leaning the top back Place the lectern by the door Wheel the Base unit into the room to its validated position If there are aeration units they must be placed in to their validated position...

Страница 15: ...ion The light ring will illuminate Ensure the white switch is not obstructed as it is the main disconnect device If the ProteQ is to communicate with the building s systems then it may be necessary to...

Страница 16: ...cycle There is sufficient liquid to run the cycle no action required Left Hand Bottle Empty expired or no bottle present Right Hand Bottle Empty expired or no bottle present Both bottles are empty ex...

Страница 17: ...handling hydrogen peroxide bottles As a minimum wear suitable gloves and goggles Press the bottom right hand side of the flap under the bottle storage compartment The flap will spring up Remove a pai...

Страница 18: ...tle with the other using the finger grips These should be facing forward for correct alignment Then slide the bottle in For 950ml bottles Pull the bottle tray open with one hand while placing the bott...

Страница 19: ...lexi Adaptor Fill the flexi adaptor with Bioquell approved Hydrogen Peroxide and place lids on both left and right hand sides and wear eye protection and gloves as a minimum when carrying it WARNING A...

Страница 20: ...the enclosure are properly sealed If large leaks occur and high concentration Hydrogen Peroxide escapes into areas where people are present it can cause serious harm Use tape provided to seal door Pre...

Страница 21: ...e Full screen navigation is given in Section 20 Start decontamination page Cycle name Quantity of peroxide required If Flexi Adapter fitted Refer to section 9 to change the cycle Expiry date of hydrog...

Страница 22: ...given below along with the light colour of the ring around the vaporiser module any installed aeration modules and which may also be repeated on the bottle module Icon Cycle stage Light colour of rin...

Страница 23: ...the button These give Enclosure temperature Enclosure relative humidity Hydrogen peroxide concentration Press to return to the cycle stage screen Press to view graphical trend WARNING Do not enter an...

Страница 24: ...rial number 6 Cycle count 7 Calibration due date 8 H2O2 lot number 9 The active programme name and parameters 10 User name 11 Signature manual input required At each transition between cycle phases th...

Страница 25: ...on can be pressed as indicated on the left To remove the printed report pull the paper up to rip off To change the paper in the printer refer to section 14 1 1 If the paper runs out the printer will p...

Страница 26: ...e the cycle recipe required by selecting it from the list and pressing To view the parameters of the cycle selected press in the bottom right of the screen The following screen will appear Parameters...

Страница 27: ...g or enclosure control system Aeration Only Cycle Check or Uncheck Check to select Aeration Only cycle Gassing Parametric Check or Uncheck Check to allow the gassing time to be automatically optimised...

Страница 28: ...to 400 Room volume to be decontaminated gassing parametric enabled Dwell Injection Rate g min 4 to 16 4 to 12 with Flexi Adaptor 4 to 10 100V H2O2 injection rate during dwell phase Dwell Time Sec 60...

Страница 29: ...rators If date and time is synchronized with the server it will be updated at the set frequency when the lectern is connected to the network 11 4 Manage Users and Passwords Administrator access only T...

Страница 30: ...ser is to be given Select only one There are three user groups that a person can be put in these are listed below OPERATOR Someone in this group can operate start and stop a pre programmed cycle trans...

Страница 31: ...ect the name of the user to be removed Select Remove A prompt box appears Select YES to remove the user or NO to return to the previous screen 11 4 4 Changing Password A local user can change their pa...

Страница 32: ...re on the hard disk Select to export the database to the hard disk and create a PDF report of it It is possible to download onto the Hard disk data logged out of cycle Navigate to the graphical view a...

Страница 33: ...ase unit 3 Wheel the Base unit out of the room to the lectern s location 4 Place the hand held low level sensor on the lectern or return it to its charging unit 5 Unplug the lectern s power cable DO N...

Страница 34: ...2 Pairing a Lectern to a ProteQ Base Unit If a ProteQ base unit requires pairing with a lectern then use the following procedure Power down and power up the lectern and log on Once paired push the op...

Страница 35: ...The following explains how to join an aeration unit to the wireless network The base unit and lectern have to already be paired Press the Open Network button The button will start to flash Then press...

Страница 36: ...The cable is plugged into the lectern and the box can then be located nearer to the ProteQ base unit allowing a better signal After plugging the repeater box in go to the administration screen and che...

Страница 37: ...oll of paper Replace with new roll of paper ensuring correct orientation as shown Push cover closed The printer is now ready to use Depressing the button will allow paper to be fed through the printer...

Страница 38: ...950ml bottles of 35 hydrogen peroxide 2Litre Bottle Part Number BQ783 XX 2000 This part number is for four 2litre bottles of 35 hydrogen peroxide where XX refers to the country region codes refer to y...

Страница 39: ...Hydrogen Peroxide If the expiry date box is yellow and flashing the hydrogen peroxide in the bottle is beyond its expiry date and cannot be used Replace bottle and repeat Cycle Run procedure Bottle R...

Страница 40: ...op the cycle otherwise will go to emergency aeration Turn the Emergency Stop button to reset when safe to enter room See Section 4 for location 2 RFID reader 1 error Left hand RFID reader alarm reads...

Страница 41: ...and contact service provider 13 Distribution fan 1 1 fault Distribution fan 1 has stopped working Will stop cycle prior to gassing stage otherwise will go into emergency aeration Aerate until room rea...

Страница 42: ...r to gassing will stop the cycle otherwise will go to emergency aeration Aerate until room reaches 1ppm and contact service provider 27 Aeration unit 1 distribution fan 1 failure if present Aeration d...

Страница 43: ...ntact service provider 37 Aeration unit 4 distribution fan 2 failure if present Aeration distribution fan has stopped working Will stop cycle prior to gassing stage otherwise will go into emergency ae...

Страница 44: ...Will prevent a cycle from starting If in cycle prior to gassing will stop the cycle otherwise will go to emergency aeration Aerate until room reaches 1ppm and contact service provider 48 Insufficient...

Страница 45: ...e prior to gassing stage otherwise will go into emergency aeration Aerate until room reaches 1ppm then re run cycle with other aeration units and contact service provider 62 Invalid Cycle Recipe Cycle...

Страница 46: ...gassing stage otherwise will go into emergency aeration Aerate until room reaches 1ppm then re run cycle with other aeration units and contact service provider 73 Insufficient Aeration Units for Load...

Страница 47: ...o Action Arrange a service visit by the Bioquell Service provider 15 3 Fault Finding Fault User Checks Further Action No power to the screen or base unit Check power is switched on at the lectern base...

Страница 48: ...category II Power Consumption Base Unit lectern 230V AC 1 8 kW max 0 2kW max 120V AC 1 8 kW max 0 2kW max 100V AC 1 8 kW max 0 2kW max Cable lengths Mains Power Base Unit 5m 16 4ft Lectern 3m 10 ft Op...

Страница 49: ...dBm Optional additional Aeration Capability 2 off onboard stored free standing wirelessly controlled units Independently powered 100 240V 3 1A Upto 6 can be operated Maximum H202 liquid injection rate...

Страница 50: ...the alarm will operate in double bursts To acknowledge the alarm press OK button this may not mute the alarm if the level is above the high alarm limit When the instrument is turned off it is still p...

Страница 51: ...ils the above is intended to be a quick reference guide 1 2 3 Press button 3 to turn the instrument ON and OFF Once ON to measure room volume press the central FUNC button 2 then the button twice 4 wh...

Страница 52: ...Biological Indicators box of 100 H20990614 Bioquell Sealing Tape H20040010 Gloves order 100 for a box of 100 H07990042 Goggles Where XX refers to the country region codes refer to your local Bioquell...

Страница 53: ...BIOQUELL PROTEQ USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Page 53 of 58 20 SCREEN NAVIGATION...

Страница 54: ...BIOQUELL PROTEQ USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Page 54 of 58...

Страница 55: ...BIOQUELL PROTEQ USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Page 55 of 58 21 DECLARATION OF CONFORMITY...

Страница 56: ...of seniority and training It is advised that this page is removed from the manual and stored in a safe place Name Password ADMINISTRATORS admin SUPERVISOR super OPERATOR oper These default passwords...

Страница 57: ...BIOQUELL PROTEQ USER MANUAL EN TD092 O M 001 REVISION 5 DRAFT Page 57 of 58 THIS PAGE IS INTENTIALLY BLANK...

Страница 58: ...O M 001 REVISION 5 DRAFT Page 58 of 58 For further information on Bioquell s technology Bioquell UK Ltd 52 Royce Close West Portway Andover Hampshire SP10 3TS T 44 0 1264 835 835 F 44 0 1264 835 836 E...

Отзывы: