77
BM3Vet Pro
User’s Manual
diagnosis.
LETHAL: Performs the detection of
Asys, VTAC/VFIB, and VTAC at the
selected lead
FULL: Performs the detection of all 13
arrhythmia.
A-1-2.
Arrhythmia Alarm
Condition
Alarm setting menu by arrhythmia type
Warning
Display Heart Beat Equipment Signal
Heart Beat equipment signal displays when the PACE mode is. The signal appears series
form. The signal size or form are meaningless clinically
Number Of Heart Beat
Attention to the animal with heart beat equipment. The heart beat equipment can show
heart beat even during arrhythmia continuously. Therefore, do not depend on heart beat
alarm excessively.
Warning
VENTRICULAR ARRHYTHMIAS
The arrhythmia analysis program is intended to detect ventricular arrhythmia. This program
is not designed to detect trial or supra ventricular arrhythmias. In some cases, it may not be
possible to distinguish the presence or absence of arrhythmias. Therefore, doctors should
analyze the arrhythmia information like other medical information.
SUSPENDED ANALYSIS
Certain conditions can delay the arrhythmia analysis. Detection and alarms associated with
arrhythmias do not occur when arrhythmia conditions are delayed. This message is
generated when the arrhythmia analysis is delayed:
LEADS FAULT, ALARM PAUSE, ALL ALARMS OFF, DISCHARGED.
Содержание BM3Vet Pro
Страница 22: ...22 BM3Vet Pro User s Manual Left Side 1 Printer 1...
Страница 45: ...45 BM3Vet Pro User s Manual Display Mode Wave and Numeric Mode Numeric Mode...
Страница 68: ...68 BM3Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 146: ...146 BM3Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 204: ...204 BM3Vet Pro User s Manual less than minus number percent plus or minus...