169
BM3Vet Pro
User’s Manual
This veterinary patient monitoring device communicates over a 2.4 GHz 802.11b / g
wireless network. Other equipment may interfere with data reception on this wireless
network. This is also true if the equipment complies with the CISPR emission
requirements. When using veterinary patient monitoring equipment to communicate
over a wireless network, be sure to check that it is compatible with existing or new
wireless systems (eg, cell phones, pager systems, cordless phones, etc.). For example, a
Bluetooth-compliant device using the 2.4 GHz frequency band may interfere with the
wireless communication of the veterinary patient monitor . For more information on
wireless deployment, please contact your Bionet representative.
Low amplitude signals such as EEG and ECG are particularly sensitive to interference
from electromagnetic energy. This equipment complies with the tests listed at the
bottom, but does not guarantee complete operation. The "quiet" electrical environment
is better. In general, the greater the distance between electrical equipment, the lower the
likelihood of interference.
M
anufacturer’s declaration
- electromagnetic emission
The BM3 Vet Pro system is intended for use in the electromagnetic environment specified below. The
customer or the user of BM3 Vet Pro system should assure that it is used in such an environment
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The BM3 Vet Pro system uses RF energy only for its
internal function. Therefore. Its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment
RF emissions
CISPR 11
Class A
The BM3 Vet Pro system is suitable for use in all es
tablishments other than domestic and those direct
ly connected to the public low-voltage power sup
plies buildings used for domestic purposes.
Harmonics emission
IEC 61000-3-2
A
Voltage fluctuation
IEC 61000-3-3
Complies
Содержание BM3Vet Pro
Страница 22: ...22 BM3Vet Pro User s Manual Left Side 1 Printer 1...
Страница 45: ...45 BM3Vet Pro User s Manual Display Mode Wave and Numeric Mode Numeric Mode...
Страница 68: ...68 BM3Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 146: ...146 BM3Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 204: ...204 BM3Vet Pro User s Manual less than minus number percent plus or minus...