Start the adapter zero. The maximum
time for a CAPNOSTAT zero is
40seconds. The typical time for a zero
is 15~20seconds.
C-1.
MODULE RESET
EtCO
2
MODULE initializing.
Note
For best result, connect the CAPNOSTAT 5 CO
2
Sensor to an adapter and wait 2 minutes
before performing the Adapter Zero procedure.
Status Message
Following is a list of some of the message that may appear on the monitor when monitoring CO
2
.
The message should clear when normal operating criteria are met or a solution is found.
* SENSOR OVER TEMP
- Cause: The sensor temperature is greater than 104
° F
- Solution: Make sure sensor is not exposed to extreme heat (heat lamp, etc.)
* SENSOR FAULTY
- Cause: One of the following conditions exist: Capnostat Source Current Failure
EEPROM Checksum Faulty, Hardware Error
- Solution: Check that the sensor is properly plugged in. Reinsert or reset the sensor if
necessary.
* SENSOR WARM UP
- Cause: Sensor under temperature, temperature not stable, Source Current unstable
- Solution: This error condition is normal at startup. This error should clear when the warmup is
complete.
* CHECK SAMPLING LINE
- Cause: This error occurs whenever the pneumatic pressure is outside the expected range.
Содержание BM3Vet Pro
Страница 22: ...22 BM3Vet Pro User s Manual Left Side 1 Printer 1...
Страница 45: ...45 BM3Vet Pro User s Manual Display Mode Wave and Numeric Mode Numeric Mode...
Страница 68: ...68 BM3Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 146: ...146 BM3Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 204: ...204 BM3Vet Pro User s Manual less than minus number percent plus or minus...