103
BM3Vet Pro
User’s Manual
①
EtCO
2
CO
2
concentration alarm upper and lower limit value display
②
Apnea alarm set time in seconds
③
Display CO
2
concentration value at exhalation
④
Display the carbon dioxide concentration value at inhalation
⑤
Show respiratory rate per minute
EtCO
2
Setup
A.
EtCO
2
menu
Menu
Description
Available settings
A-1.
ALARM
EtCO
2
Alarm Setup Menu
A-1-1.
PARAMETER ALARM
LIMIT
ETCO
2
, FICO2, AWRR, APNEA
parameter alarm, level, action setup
menu
A-1-2.
TECHNICAL ALARM
CONDITION
ETCO2-MODULE OFF
ETCO2-CHECK ADAPTOR
ETCO2-CHECK LINE
ETCO2-CHECKLINEDISCONNECT
ETCO2-CO2INVALID
ETCO2-OVERRANGE
ETCO2-ZEROREQUIRED
ETCO2-SYSTEMFAULT
ETCO2-TEMPUNSTABLE
A-2.
DISPLAY OPTION
EtCO
2
Parameter wave display setup
menu.