148
BM3Vet Pro
User’s Manual
Feature
Specification
CO2
- Range
0
–
100 mmHg; 0
–
13.3 kPa; 0-10% CO2 STPD (standard
temperature and pressure dry)
- Accuracy
± (0.2 vol% + 4% relative)
- Rise Time
400 ms (average)
Anesthetic Agents
- Gases
Isoflurane, Enflurane, Halothane, Desflurane, Sevoflurane
- Range
Iso., Enf., Hal., Sev.: 0
–
6%
Des.: 0
–
18%
- Resolution
0 .01%
- Accuracy
± (0.15% vol% + 4% relative)
- Rise Time
450 ms (average)
Calibration
Factory calibrated
Power Up Time
30 sec
Delay Time
< 4 sec
Dimensions
175mm x 85mm x 50mm (L x W x H)
Respiration
- Range
0-150 breaths/min
- Accuracy
±1 breath/min
Flow Rate
- Range
170 ml/min
- Accuracy
±20ml/min
Environment Condition
- Operating Temperature
15
–
35°C
- Storage Temperature
-
5 to 50°C
- Ambient Humidity
15-95% RH
- Ambient Pressure
70kPa to 106kPa (525mmHg to 795mmHg)
Weight
400g
Water Removal System
Water trap tank
Sampling Line
Anesthetic gas tolerant standard sample line
Appication
Veterinary use Only
Species: Canine, Feline, Equine
Содержание BM3Vet Pro
Страница 22: ...22 BM3Vet Pro User s Manual Left Side 1 Printer 1...
Страница 45: ...45 BM3Vet Pro User s Manual Display Mode Wave and Numeric Mode Numeric Mode...
Страница 68: ...68 BM3Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 146: ...146 BM3Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 204: ...204 BM3Vet Pro User s Manual less than minus number percent plus or minus...