133
BM3Vet Pro
User’s Manual
Trouble Shooting and Solutions
When the measurement density error is high during operation
①
Check the connection of sampling line and water trap and reconnect.
②
Check the connection between the water trap and the water trap holder and reconnect them.
③
Perform manual calibration in an unpolluted environment.
④
If you close the gas input part of the water trap by hand and operate it for a few seconds,
“Occlusion”
will appear on the host device. At this time, the motor operates strongly and absorbs
gas from the gas input part. If the above operation does not appear, please contact our head
office for inspection.
Regular inspection
As with all medical equipment, we encourage you to have your device tested regularly once a
year for BMGA.
Содержание BM3Vet Pro
Страница 22: ...22 BM3Vet Pro User s Manual Left Side 1 Printer 1...
Страница 45: ...45 BM3Vet Pro User s Manual Display Mode Wave and Numeric Mode Numeric Mode...
Страница 68: ...68 BM3Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 146: ...146 BM3Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 204: ...204 BM3Vet Pro User s Manual less than minus number percent plus or minus...