8. If the cuvette pressure value changes more, troubleshoot tubing and fittings. If the
problem persists, return the analyzer to PHASEIN.
Note:
In step 5, if the cuvette pressure is less than 15 kPa below atmospheric
pressure, repeat steps 1 to 3 blocking the exhaust port quicker.
Gas span check
Confirm the gas measurements values by performing the following sequence
1. Connect a new Nomoline sampling line to the ISA gas analyzer.
2. Warm up for at least 1 min.
3. Connect the calibration gas to the ISA gas analyzer. The calibration gas shall have
an accuracy of at least ±
0.03 vol% or ±
(0.02 vol% +0.1% of reading), whichever
is greater.
4. Supply the calibration gas.
5. Check that the displayed gas concentration readings correspond to the calibration
gas values.
6. The gas readings shall be within the following ranges:
a. CO
2
± (
0.3 vol% + 4% of reading)
b. N
2
O
± (
2 vol% + 5% of reading)
c. Agents ±
(0.2 vol% + 10% of reading)
d. O
2
± (
2 vol% + 2% of reading)
7. If the gas readings are outside the specified range, perform a gas span calibration
for the failing gases. If failing on Agents, the span calibration shall always be
performed using DES.
Gas span calibration
Only
perform the gas span calibration if the gas span check fails repeatedly.
Содержание BM3Vet Pro
Страница 22: ...22 BM3Vet Pro User s Manual Left Side 1 Printer 1...
Страница 45: ...45 BM3Vet Pro User s Manual Display Mode Wave and Numeric Mode Numeric Mode...
Страница 68: ...68 BM3Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 146: ...146 BM3Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 204: ...204 BM3Vet Pro User s Manual less than minus number percent plus or minus...