Step 7:
Using the 3mm Allen wrench provided in the accessories kit,
unscrew the three screws mounting the lamp assembly to the lamp
housing.
CAUTION! Fingerprints can ruin the FL600 quartz halogen
lamp and render it unusable.
Step 8:
Disconnect the lamp wires, and slide the lamp assembly away from
the PMT base and toward the fan.
VROHQRLG
FDEOH
FDEOH KDUQHVV
ZLUHV
FDEOH KDUQHVV
307 EDVH
Figure 4-6
Step 9:
Carefully remove the lamp assembly from the lamp housing.
Step 10:
Remove the lamp from the lamp assembly by removing the three
3mm hex screws.
Step 11:
Insert the new lamp (P/N 6000556S) into the lamp assembly, and
then replace the three screws.
Step 12:
Remount the lamp assembly to the lamp housing, and screw down
the three 3mm hex screws. Be careful not to over-tighten them.
Step 13:
Reconnect the two lamp wires in any configuration. Polarity is not
an issue.
Step 14:
Replace the lamp cover, making sure that cable attached to the
solenoid is routed through the center hole of the lamp cover.