Bio-Tek FL600
3-5
y
An emission filter has an “M” designation.
•
A dual-function Excitation/Emission filter has an
“X/M“ designation.
Absorbance Filters:
Insert the filter into the filter holder, and insert this
assembly into the filter wheel.
Absorbance filters do not need to be aligned.
Step 8:
Note the filter’s position on the wheel (the molded letters A, B, C
are to the right of each filter), and the wavelength that is printed on
the top or side of each filter cell.
11\Y\YW^
W^]U
]U^^dd
11bbb_g
b_g
Figure 3-5: Aligning the filter with the filter wheel
Step 9:
Replace the covers on the filter wheels. Put any unused filters in a
secure place.
Secure the filter wheel covers by gently placing the cover
over the access hole and then pressing down on the plastic
snap covers.
Fit the instrument cover, and secure it using the five screws.
Seal the unused filters in a plastic container, and store them
in a cool, dry location.
Step 10:
Re-connect the power cable, and power up the reader.
The plate carrier will move to the plate load position.