5-4
Technical Information
7HPSHUDWXUH&RQWURO
The FL600 is fitted with a heater block which can be programmed to heat the
plate to a temperature above ambient.
Temperature Range
A6
°
Celsius to 50
°
Celsius
Uniformity
Within ±0.5
°
Celsius across a 96 well Dynatech black plate
measured with thermal probes in the center and four corner
wells and 150
µ
l of Johnsons Baby Oil per well.
Time to Stabilize
Within 30 minutes after switching on
Time to Cool
Approx. 2.5 hours from 37
°
Celsius to 6
°
Celsius
above the on-board ambient temperature reference
sensor.
3HUIRUPDQFH6SHFLILFDWLRQV
Standard
Plate Types
All 96-well microplates and most 48- to 6-well
tissue culture plates
High-Density Plate Types
384-well microplates.
Note: Reader performance will be optimized by factory-
fitting of a smaller fibre optic probe 1.0mm or 1.5mm
diameter).
60, 72 and 96 well Terasaki plates
Note: Optimum probe size is 1.0mm or 1.5mm diameter.
A custom plate adapter is required.
Plate Sizes
Minimum 126 mm
×
84 mm
Maximum 133 mm
×
90 mm