Technical Information
5-11
)/6HULDO&RPPXQLFDWLRQV
The reader communicates with the computer via a 9-pin D-type male-female
connector (P/N 75034).
The connector is wired as follows.
PIN
Computer
FL600
1
DCD
Not used
2
RX
TX data to computer
3
TX
RX data from
computer
4
DTR
Not used
5
GND
Signal ground
6
DSR
Not used
7
RTS
Linked to pin 8
8
CTS
Linked to pin 7
9
RI
Not used
Serial Port
Parameter
KC4 Reader
Control and Data
Reduction
Software
FL600
Diagnostic
Software
Baud rate
9600
9600
Data bits
8
8
Parity
Even
None
Stop bit
1
1
Note: The FL600 Diagnostic Software settings can not be changed- through
either software or hardware tools.