Technical Information
5-5
Sensitivity (Specified)
Bottom Reading 5.0mm Probe
10 pg/ml solution of Sodium Fluoroscein, in PBS,
150
µ
l per well Signal to Noise Ratio of 2 or greater.
Excitation 485/20, Emission 530/25
Hellma 96-well quartz plate.
125 ng/ml solution of Propidium Iodide, in PBS,
150
µ
l per well Signal to Noise Ratio of 2 or greater.
Excitation 485/20, Emission 645/40
Cestar 3603 96-well plate.
Top Reading 5.0mm Probe
20 pg/ml solution of Sodium Fluoroscein, in PBS,
300
µ
l per well Signal to Noise Ratio of 2 or greater.
Excitation 485/20, Emission 530/25
Dynatech 96-well black plate.
160 pg/ml solution of Methylumbelliferone, in CBB,
300
µ
l per well Signal to Noise Ratio of 2 or greater.
Excitation 360/40, Emission 460/40
Dynatech 96-well black plate.
125 ng/ml solution of Propidium Iodide, in PBS,
300
µ
l per well Signal to Noise Ratio of 2 or greater.
Excitation 485/20, Emission 645/40
Dynatech 96-well black plate.
Sensitivity (Typical)
Typical sensitivities can be two to three times
more sensitive than specified