KBF P + KBF LQC + KBWF (E6) 11/2016
page 46/189
5.6
Principle of controller entries
In the selection and entry menus there are icons displayed in the footers which you can use to take over
the entry or cancel it.
Selection menu (example)
Entry menu (example)
There are the following possibilities:
Press the
Confirm
icon to take over the entries and exit the menu or continue the menu se-
quence.
Press the
Close
icon to exit the menu or cancel the menu sequence without taking over the en-
tries.
When terminating a menu sequence, an information window appears, which must be confirmed.
5.7
Performance during and after power failures
During a power failure, all controller functions are shut down. The zero-voltage relay alarm output (option,
chap. 21.5) is switched to alarm position for the whole duration of the power failure.
After the power returns, all functions return to the same status the chamber had before power failure. .
The controller continues to function in the original operating mode it was in previously before the power
failure occurred.
•
Performance after power failure in Idle mode
Control is deactivated
•
Performance after power failure in Fixed value operation mode
All functions return to the same status the chamber had before power failure. The set-points are im-
mediately resumed.
•
Performance after power failure during time program operation
The program is resumed at the point where the interruption occurred with the latest set-points reached
during the program run.
•
Performance after power failure during week program operation
The week program continues with the values corresponding to the current time.
Power failure and power return are noted in the event list (chap. 16.3).
If during power failure an alarm has occurred (tolerance range, safety controller, temperature safety de-
vice class 3.3 (option), confirm the alarm. See chap. 12.3.
KBF LQC:
Exposition with radiation is continued in the same was as it had been set manually or automati-
cally before power failure. Integration of the light values continues with the dose values reached before
power failure.
Содержание KBF P 240-UL
Страница 173: ...KBF P KBF LQC KBWF E6 11 2016 page 173 189 25 7 Dimensions size 240...
Страница 174: ...KBF P KBF LQC KBWF E6 11 2016 page 174 189 25 8 Dimensions size 720...
Страница 176: ...KBF P KBF LQC KBWF E6 11 2016 page 176 189...
Страница 177: ...KBF P KBF LQC KBWF E6 11 2016 page 177 189 26 2 EU Declaration of Conformity for KBF LQC...
Страница 178: ...KBF P KBF LQC KBWF E6 11 2016 page 178 189...
Страница 179: ...KBF P KBF LQC KBWF E6 11 2016 page 179 189 26 3 EU Declaration of Conformity for KBWF...
Страница 180: ...KBF P KBF LQC KBWF E6 11 2016 page 180 189...
Страница 182: ...KBF P KBF LQC KBWF E6 11 2016 page 182 189...
Страница 183: ...KBF P KBF LQC KBWF E6 11 2016 page 183 189 27 Product registration...