4. Installation and connections
4.1
Spacer for wall distance
Please fix both spacers with the supplied screws at the chamber rear. This serves to ensure the pre-
scribed minimum distance to the rear wall of 100 mm /
3.94
in
.
Figure 8: Spacer for wall distance
Figure 9: Rear KMF with mounted spacers
4.2
Wastewater connection
Fasten the wastewater hose to the wastewater connection “OUT” (14) on the rear of the chamber (olive
∅
14 mm). Observe the following points:
•
You can use a part of the supplied water hose as a drainage hose. In case another hose is used, it has
to be permanently resistant against at least 95 °C /
203 °F
.
•
Mount the wastewater hose with a maximum positive inclination of 1 m and a maximum total length of
3 m.
•
Protect the chamber end of the drainage hose with one of the supplied hose clamps.
•
Reliably prevent sucking back of wastewater. The end of the wastewater hose must not be immersed
in liquids. This can be ensured e.g., by free discharge.
Waste water is collected in an internal can with a volume of approx. 0.5 liters. It is pumped off
only when required, thus there is no continuous waste water flow.
Protect the waste water supply with the supplied hose clamps.
Содержание KBF P 240-UL
Страница 173: ...KBF P KBF LQC KBWF E6 11 2016 page 173 189 25 7 Dimensions size 240...
Страница 174: ...KBF P KBF LQC KBWF E6 11 2016 page 174 189 25 8 Dimensions size 720...
Страница 176: ...KBF P KBF LQC KBWF E6 11 2016 page 176 189...
Страница 177: ...KBF P KBF LQC KBWF E6 11 2016 page 177 189 26 2 EU Declaration of Conformity for KBF LQC...
Страница 178: ...KBF P KBF LQC KBWF E6 11 2016 page 178 189...
Страница 179: ...KBF P KBF LQC KBWF E6 11 2016 page 179 189 26 3 EU Declaration of Conformity for KBWF...
Страница 180: ...KBF P KBF LQC KBWF E6 11 2016 page 180 189...
Страница 182: ...KBF P KBF LQC KBWF E6 11 2016 page 182 189...
Страница 183: ...KBF P KBF LQC KBWF E6 11 2016 page 183 189 27 Product registration...