KBF P + KBF LQC + KBWF (E6) 11/2016
page 33/189
4.6
Electrical connection
The chambers are supplied ready for connection. They come with a fixed power connection cable of at
least 1800 mm /
70.87 in
in length.
Model version
Art. No.
(x = 0 or 1)
Power plug
V/-10% at the
indicated power fre-
quency
Current
type
Chamber
fuse
KBWF 240
KBF P 240
KBF LQC 240
9x20-0336
9x20-0328
9x20-0332
Shock-proof plug
200-230 V at 50 Hz
1N~
16 Amp
KBF P 240-UL
KBF LQC 240-UL
9x20-0329
9x20-0333
NEMA 6-20P
200-240 V at 50Hz
200-240 V at 60Hz
2~
16 Amp
KBWF 720
KBF P 720
KBF LQC 720
9x20-0337
9x20-0330
9x20-0334
Shock-proof plug
200-230 V at 50 Hz
1N~
16 Amp
KBF P 720-UL
KBF LQC 720-UL
9x20-0331
9x20-0335
NEMA 6-20P
200-240 V at 50Hz
200-240 V at 60Hz
2~
16 Amp
•
The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the chamber’s protective conductor meets the lat-
est technology. The protective conductors of the socket and plug must be compatible!
•
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the chamber’s type plate (left chamber side, bottom right-hand, see chap. 1.4).
•
When connecting, please observe the regulations specified by the local electricity supply company and
as well as the VDE directives (for Germany). We recommend the use of a residual current circuit
breaker.
•
Pollution degree (acc. to IEC 61010-1): 2
•
Installation category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 25.4).
To completely separate the chamber from the power supply, you must disconnect the power
plug. Install the chamber in a way that the power plug is easily accessible and can be easily
pulled in case of danger.
Remark when operating the chamber with a power frequency of 60 Hz:
WARNING
High leakage current.
Electrical hazard.
Earth connection essential before connecting supply. Check socket before inserting
plug.
Содержание KBF P 240-UL
Страница 173: ...KBF P KBF LQC KBWF E6 11 2016 page 173 189 25 7 Dimensions size 240...
Страница 174: ...KBF P KBF LQC KBWF E6 11 2016 page 174 189 25 8 Dimensions size 720...
Страница 176: ...KBF P KBF LQC KBWF E6 11 2016 page 176 189...
Страница 177: ...KBF P KBF LQC KBWF E6 11 2016 page 177 189 26 2 EU Declaration of Conformity for KBF LQC...
Страница 178: ...KBF P KBF LQC KBWF E6 11 2016 page 178 189...
Страница 179: ...KBF P KBF LQC KBWF E6 11 2016 page 179 189 26 3 EU Declaration of Conformity for KBWF...
Страница 180: ...KBF P KBF LQC KBWF E6 11 2016 page 180 189...
Страница 182: ...KBF P KBF LQC KBWF E6 11 2016 page 182 189...
Страница 183: ...KBF P KBF LQC KBWF E6 11 2016 page 183 189 27 Product registration...