KBF P + KBF LQC + KBWF (E6) 11/2016
page 142/189
This position is only meant for the light cassettes. Due to their connections, shelves cannot
be positioned here.
You can move the light cassettes to the indicated positions.
Plug
rear
front
Figure 26: Arrangement of the fluorescent tubes in the light cassette of KBF P / KBF LQC
Plug
rear
front
Figure 27: Arrangement of the fluorescent tubes in the light cassette of KBWF
You will obtain optimum homogeneity by alternately placing the fluorescent tubes of the same type, i.e.,
opposite arrangement of the inscription:
Inscription
Figure 28: Opposite arrangement of two fluorescent tubes
When replacing the fluorescent tubes, observe tube orientation (inscription).
How to replace the fluorescent tubes is described in chap.22.2.
Operation with light cassettes and illumination on: Maximum temperature 60 °C /
140 °F
.
Operation with light cassettes and illumination off: Do also NOT operate the chamber at tem-
peratures >60 °C /
140 °F.
Otherwise, the lifetime of the fluorescent tubes will be considerably
reduced.
When operating the chamber at temperatures > 60 °C /
140 °F
, remove the light cassettes.
Содержание KBF P 240-UL
Страница 173: ...KBF P KBF LQC KBWF E6 11 2016 page 173 189 25 7 Dimensions size 240...
Страница 174: ...KBF P KBF LQC KBWF E6 11 2016 page 174 189 25 8 Dimensions size 720...
Страница 176: ...KBF P KBF LQC KBWF E6 11 2016 page 176 189...
Страница 177: ...KBF P KBF LQC KBWF E6 11 2016 page 177 189 26 2 EU Declaration of Conformity for KBF LQC...
Страница 178: ...KBF P KBF LQC KBWF E6 11 2016 page 178 189...
Страница 179: ...KBF P KBF LQC KBWF E6 11 2016 page 179 189 26 3 EU Declaration of Conformity for KBWF...
Страница 180: ...KBF P KBF LQC KBWF E6 11 2016 page 180 189...
Страница 182: ...KBF P KBF LQC KBWF E6 11 2016 page 182 189...
Страница 183: ...KBF P KBF LQC KBWF E6 11 2016 page 183 189 27 Product registration...