KBF P + KBF LQC + KBWF (E6) 11/2016
page 27/189
4.3
Freshwater supply
Connect the waste water pipe before connecting the chamber to a freshwater pipe or filling the
freshwater can (option, chap. 21.9).
You can supply the chamber with freshwater via a water pipe or by manually filling a freshwater can (op-
tion, chap. 21.9).
Water intake temperature NOT below +5 °C /
41 °F
and not exceeding 40 °C /
104 °F
.
CAUTION
Calcification of the humidifying system.
Damage to the chamber.
Operate the chamber with deionized (demineralized) water only.
Types of suitable water quality:
•
Deionized water from a water treatment installation already existing at the customer's site. Conductivity
from 1 µS /cm up to a maximum of 20 µS/cm. (Water, which is in equilibrium with the CO
2
in the air,
and has a conductivity below 1 µS/cm (ultrapure water), may cause acid corrosion due to its low pH.)
•
Water treated by the optional water treatment system BINDER Pure Aqua Service (disposable system).
A reusable measuring equipment to assess the water quality is included (chap. 21.10).
BINDER GmbH is NOT responsible for the water quality at the user’s site.
Any problems and malfunctions that might arise following use of water of deviating quality is
excluded from liability by BINDER GmbH.
The warranty becomes void in the event of use of water of deviating quality.
4.3.1 Automatic fresh water supply via water pipe
An enclosure inside the chamber contains the connection kit for freshwater and wastewater. Install the
freshwater connection using either the enclosed water hose or another pressure-resistant one. To accom-
plish this, remove the cover of the freshwater connection “IN” (13) on the rear of the chamber. Protect
both ends of the hose with two of the four supplied hose clamps.
Before turning on the chamber, check the connection for leaks. Water supply is automatically effected via
the freshwater connection “IN” (13).
As the appliance only lets in water when required, there is no continuous water flow.
•
Supply pressure 1 to 10 bar when connecting to a water pipe
•
Water type: deionized (demineralized) water
•
Water intake temperature NOT below +5 °C /
41 °F
and not exceeding 40 °C /
104 °F
.
•
The water intake should be provided with a shut-off slide or water-tap.
•
For the water supply, fix the delivered adapter with hose olive on the thread at the rear of
the chamber.
•
Protect the water supply at one side with the supplied hose clamp.
Содержание KBF P 240-UL
Страница 173: ...KBF P KBF LQC KBWF E6 11 2016 page 173 189 25 7 Dimensions size 240...
Страница 174: ...KBF P KBF LQC KBWF E6 11 2016 page 174 189 25 8 Dimensions size 720...
Страница 176: ...KBF P KBF LQC KBWF E6 11 2016 page 176 189...
Страница 177: ...KBF P KBF LQC KBWF E6 11 2016 page 177 189 26 2 EU Declaration of Conformity for KBF LQC...
Страница 178: ...KBF P KBF LQC KBWF E6 11 2016 page 178 189...
Страница 179: ...KBF P KBF LQC KBWF E6 11 2016 page 179 189 26 3 EU Declaration of Conformity for KBWF...
Страница 180: ...KBF P KBF LQC KBWF E6 11 2016 page 180 189...
Страница 182: ...KBF P KBF LQC KBWF E6 11 2016 page 182 189...
Страница 183: ...KBF P KBF LQC KBWF E6 11 2016 page 183 189 27 Product registration...