Alarm message
Start after condition occurred
“Temperature range”
after configurable time
“CO2 range
after configurable time
“O2 range
after configurable time
“Humidity range
after configurable time
“Door open”
after configurable time
If the condition (deviation from the tolerance range or door opening) ends within the configured alarm
delay time, no alarm will be triggered after the delay time.
When operating the chamber with deactivated CO
2
, O
2
(chamber with O
2
control) or humidity
(chamber with active humidification) control (chap. 6.8), the corresponding tolerance range
alarms will not occur.
Setting:
•
Select the field ”Temperature range” and enter the desired value for the temperature range. Setting
range: 1 °C to 10 °C. Confirm entry with
Confirm
icon.
•
Select the field “Temp alarm delay” and enter the time in minutes, after which the range alarm shall be
triggered. Setting range: 1 min to 120 min. Confirm entry with
Confirm
icon.
•
Select the field ”CO2 range” and enter the desired value for the CO
2
range. Setting range: 1 vol.-% to
10 vol.-%. Confirm entry with
Confirm
icon.
•
Select the field “CO2 alarm delay” and enter the time in minutes, after which the range alarm shall be
triggered. Setting range: 1 min to 120 min. Confirm entry with
Confirm
icon.
•
Select the field ”O2 range” and enter the desired value for the O
2
range. Setting range: 1 vol.-% to 10
vol.-%. Confirm entry with
Confirm
icon.
•
Select the field “O2 alarm delay” and enter the time in minutes, after which the range alarm shall be
triggered. Setting range: 1 min to 120 min. Confirm entry with
Confirm
icon.
•
Select the field “Humidity range” and enter the desired value for the humidity range. Setting range: 5%
r.h. to 20% r.h. Confirm entry with
Confirm
icon.
•
Select the field “Humid alarm delay” and enter the time in minutes, after which the range alarm shall
be triggered. Setting range: 1 min to 120 min. Confirm entry with
Confirm
icon.
•
Select the field “Door alarm delay” and enter the time in minutes, after which the range alarm shall be
triggered. Setting range: 1 min to 120 min. Confirm entry with
Confirm
icon.
After completing the settings, press the
Confirm
icon to take over the entries and exit the menu,
or
press
the
Close
icon to exit the menu without taking over the entries.