CB (E7) 06/2018
Page 89/174
Condition
Alarm message
Start after
condition occurred
Zero-voltage relay
alarm output
Exceeded setpoint of the safety
controller class 3.1
“Safety controller
immediately
----
Low CO
2
outlet pressure
(< 0.3 bar /
4.4 psi
)
“CO2 pressure”
immediately
----
Chamber with O
2
control:
Low O
2
outlet pressure
(< 0.3 bar /
4.4 psi
)
“O2 pressure”
immediately
----
Chamber with O
2
control:
Low N
2
outlet pressure
(< 0.3 bar /
4.4 psi
)
“N2 pressure”
immediately
----
Temperature sensor defective
e.g. “- - - - “ or
“<-<-< “ or “>->->”
immediately
immediately
Safety controller temperature sensor
defective
Safety controller
sensor
immediately
immediately
CO
2
sensor defective. Contact
BINDER Service.
“CO2 sensor fault”
immediately
immediately
Chamber with O
2
control:
O
2
sensor defective. Contact BINDER
Service
“O2 sensor fault”
immediately
immediately
Chamber with active humidification:
Humidity sensor defective. Contact
BINDER Service
“Humid sensor fault”
immediately
immediately
Hot air sterilization running
“Sterilization”
immediately
----
Alarm messages are displayed in the list of active alarms until acknowledging them. They are also shown
in the event list.
When operating the chamber without CO
2
/ O
2
/ N
2
connection, turn off the corresponding
gas control in the “setpoints” menu (chap. 6.8) in order to avoid tolerance range and pressure
alarms.
Chamber with active humidification:
When operating the chamber without water supply, turn
off humidity control in the “setpoints” menu (chap. 6.8) in order to avoid humidity alarms..