10.4 Operating modes
The MB2 program controller operates in the following operating modes:
•
Idle mode
The controller is not functional, i.e., there is no heating or refrigeration. The fan is off. The chamber
approximates ambient values.
You can activate and deactivate this operating mode with the “Idle mode” control contact in Fixed value
operating mode (chap. 12.3), time program operation (chap. 17.7.3) and week program operation
•
Fixed value operating mode
The controller operates as a fixed-point controller, i.e., set-points can be defined, which are then main-
tained until the next manual change (chap. 12.1).
•
Timer program operation
Stopwatch function: during an entered duration the controller constantly equilibrates to the setpoints
entered in Fixed value operation mode.
•
Time program operation
An entered time program is running. The controller offers 25 program memory places with 100 program
sections each. The total number of program sections of all programs is unlimited
•
Week program operation
An entered week program is running. The controller offers 5 program memory places with 100 switching
points each. The switching points can be distributed over all days of the week.
Содержание 9020-0402
Страница 159: ...MK LIT E5 01 2021 page 159 172 30 7 Dimensions LIT MK 240 dimensions mm...
Страница 160: ...MK LIT E5 01 2021 page 160 172 LIT MK 720 dimensions mm...
Страница 161: ...MK LIT E5 01 2021 page 161 172 31 Certificates and declarations of conformity 31 1 EU Declaration of Conformity...
Страница 162: ...MK LIT E5 01 2021 page 162 172...
Страница 163: ...MK LIT E5 01 2021 page 163 172...
Страница 165: ...MK LIT E5 01 2021 page 165 172...
Страница 166: ...MK LIT E5 01 2021 page 166 172 31 3 Safety concept certificate from T V S d...
Страница 173: ......