MK LIT (E5) 01/2021
page 73/172
12.2 Direct setpoint entry via Normal display
Alternatively you can also enter the setpoints directly via Normal display.
Normal display.
Select the setpoint you want to change.
Example: “Temperature” entry menu.
Enter the desired setpoint and confirm entry
with
Confirm
icon
12.3 Special controller functions via operation lines
Press the
Setpoint setting
icon to access the “Setpoint” setting menu from Normal display.
You can define the switching state of up to 16 operation lines (control contacts). They are used to activate
/ deactivate special controller functions.
•
Operation line “Idle mode” activates / deactivates the operating mode “Idle mode”.
•
Operation lines “Switching output 1” up to “4” can be used to turn on and off any equipment connected
to the zero-voltage switching outputs (DIN sockets (17) and (18) (option, chap. 25).
•
Operation line “Compressed air valve” serves to open the solenoid valve of the compressed air connec-
tion (for optional compressed air connection, chap. 26.5)
•
Operation line “Anti-condensation” serves to switch the anti-condensation protection (chap. 24).
The other operation lines are without function.
Содержание 9020-0402
Страница 159: ...MK LIT E5 01 2021 page 159 172 30 7 Dimensions LIT MK 240 dimensions mm...
Страница 160: ...MK LIT E5 01 2021 page 160 172 LIT MK 720 dimensions mm...
Страница 161: ...MK LIT E5 01 2021 page 161 172 31 Certificates and declarations of conformity 31 1 EU Declaration of Conformity...
Страница 162: ...MK LIT E5 01 2021 page 162 172...
Страница 163: ...MK LIT E5 01 2021 page 163 172...
Страница 165: ...MK LIT E5 01 2021 page 165 172...
Страница 166: ...MK LIT E5 01 2021 page 166 172 31 3 Safety concept certificate from T V S d...
Страница 173: ......