NOTICE
Danger of overheating due to lack of ventilation.
Damage to the voltage and frequency changer.
∅
Do NOT install the voltage and frequency changer in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
The voltage and frequency changer is equipped with
four castors. The rear castors can be easily locked
via the attached brakes
9.4 Electrical connection
9.4.1 Information on connecting the battery test chamber
The chambers are supplied ready for connection.
The chambers come with a fixed power connection cable of at least 1800 mm /
70.87 in
in length. They are
equipped with three internal overload releases against excess-current.
Model
Power plug
Nominal v/- 10% at the
indicated power frequency
Current type Chamber fuse
LIT MK 115
LIT MK 240
LIT MK 720
CEE plug 5-poles
16 Amp
400 V at 50 Hz
3 N~
16 Amp
3 x internal
•
The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the chamber’s protective conductor meets the latest
technology. The protective conductors of the socket and plug must be compatible!
DANGER
Electrical hazard due to missing protective conductor connection.
Deadly electric shock.
Make sure that the chamber’s power plug and the power socket match and se-
curely connect the electrical protective conductors of the chamber and the house
installation.
•
Only use original connection cables from BINDER according to the above specification.
•
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the chamber’s type plate (left chamber side, bottom right-hand, see chap. 1.6)
NOTICE
Danger of incorrect power supply voltage due to improper connection.
Damage to the chamber.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
•
When connecting, please observe the regulations specified by the local electricity supply company as
well as the local or national electrical regulations (VDE directives for Germany).
•
Observe a sufficient current protection according to the number of devices that you want to operate. We
recommend the use of a residual current circuit breaker.
Содержание 9020-0402
Страница 159: ...MK LIT E5 01 2021 page 159 172 30 7 Dimensions LIT MK 240 dimensions mm...
Страница 160: ...MK LIT E5 01 2021 page 160 172 LIT MK 720 dimensions mm...
Страница 161: ...MK LIT E5 01 2021 page 161 172 31 Certificates and declarations of conformity 31 1 EU Declaration of Conformity...
Страница 162: ...MK LIT E5 01 2021 page 162 172...
Страница 163: ...MK LIT E5 01 2021 page 163 172...
Страница 165: ...MK LIT E5 01 2021 page 165 172...
Страница 166: ...MK LIT E5 01 2021 page 166 172 31 3 Safety concept certificate from T V S d...
Страница 173: ......