MK LIT (E5) 01/2021
page 22/172
•
The operator is responsible for compliance with the relevant regulations on the safety and instruction of
personnel involved in the installation, installation, commissioning, operation, cleaning and decommis-
sioning as well as maintenance and repair of the chamber as well as environmental protection.
•
The battery test chamber detects the beginnings of an EUCAR Hazard levels 4 to 6 event and provides
signals via a two-stage alarm, the evaluation of which serves the user to take safety or countermeasures.
•
The operator must operate the battery test chamber in an environment that reduces all effects of a non-
containment event of EUCAR Hazard levels 4 to 6 to a safe level. Danger to persons or the environment
must be safely ruled out
.
•
The battery test chamber can be completely destroyed in the event of this event.
2.5 Medical devices
The chambers are not classified as medical devices as defined by the Medical Device Directive 93/42/EEC
and Regulation (EU) No 2017/745.
Due to the special demands of the Medical Device Directive (MDD), these chambers are not
qualified to perform sterilization of medical devices as defined by the directive 93/42/EWG.
2.6 Personnel Requirements
Only trained personnel with knowledge of the handling of hazardous substances and lithium-ion accumu-
lators as well as knowledge of the operating instructions may set up and install, commission, operate, clean
and decommission the chamber. Further technical requirements (e.g. electrical knowledge) and knowledge
of the service manual are required for maintenance and repairs. The requirements specified by the operator
for PPE (personal protective equipment) must be observed. Training on handling compressed gas cylinders
is required to replace the CO
2
compressed gas cylinder.
2.7 Installation site requirements
and ambient conditions
The chambers are designed for setting up inside a building (indoor use).
The requirements described in the Operating Manual for installation site and ambient conditions (chap. 8.4)
must be met.
Dangerous gases must be safely prevented from spreading to the surroundings of the chamber. Active
extraction (technical ventilation in accordance with national regulations) must be connected to the exhaust
port on the chamber. Extraction must remain effective throughout operation of the chamber. If the extraction
fails, the chamber must be disconnected from the power supply.
We recommend the installation of a CO
2
alarm system at the place of installation.
It must be possible to switch the chamber off immediately in the event of an error.
The chamber must not be operated without continuous monitoring.
Содержание 9020-0402
Страница 159: ...MK LIT E5 01 2021 page 159 172 30 7 Dimensions LIT MK 240 dimensions mm...
Страница 160: ...MK LIT E5 01 2021 page 160 172 LIT MK 720 dimensions mm...
Страница 161: ...MK LIT E5 01 2021 page 161 172 31 Certificates and declarations of conformity 31 1 EU Declaration of Conformity...
Страница 162: ...MK LIT E5 01 2021 page 162 172...
Страница 163: ...MK LIT E5 01 2021 page 163 172...
Страница 165: ...MK LIT E5 01 2021 page 165 172...
Страница 166: ...MK LIT E5 01 2021 page 166 172 31 3 Safety concept certificate from T V S d...
Страница 173: ......