P
ag
e
66
select “Remain”.
Once you see this display, we are now ready to move to the next step.
Remember, if you selected “remove,” you will need to remove the trial weight from
the trial mass measurement.
Now we will remove the trial weight and then place 749 mgrams at 293.6 degrees.
We can also select Component which will display a new screen. Enter the
number of components i.e., 12 components and then the balance program will
automatically calculate the component requirements between two components, or
you can select manual and choose between two components of your choice.
Tools:
Add/ Cut P1
: Changes the sign of the
mass calculation.
Tolerance
:
Adjust the
tolerance during
the run, brings up
the tolerance
display.
Unequal radii:
Sometimes
where the trial
weight is applied
and the diameter of the rotor are two different radius.
The balance program allows the user to enter the
diameter for each for correct balance calculations.
Drill Depth
: When you are removing
material for your balance process using a
drilling process, you will need to tell the
program the diameter of the drill bit and the
type of material.
Vibration History
: This allows you to
move the cursor to each point to see the
Содержание Fieldpaq
Страница 5: ...Page 4 ...
Страница 40: ...Page 39 Legend Select Off Channel No or Channel ID BG Color Select either White or Black background color ...
Страница 54: ...Page 53 Orbit x y ...
Страница 57: ...Page 56 FFT Examples 1 Insertion spike measurements 2 Fieldpaq modal test 3 Sound measurement 4 ...
Страница 61: ...Page 60 ...
Страница 104: ...Page 103 Power Spectrum Time Waveform Bearing Spectrum ...
Страница 199: ...Page 198 ...
Страница 200: ...Page 199 DVF specifications Maximum frequency band Maximum resolution Maximum file size ...
Страница 204: ...Page 203 Done ...
Страница 206: ...Page 205 5 Start the Data Explorer Program 6 Select File Import ...
Страница 215: ...Page 214 ...
Страница 222: ...6 ...