P
ag
e
131
Soft Key Menu: Tools
Selecting tools enables you to perform an IEPE check, calibrationand adding or
clearing traces to the trace plot. With linear Order Trace (peak) displayed you also
can choose Mark Cursor, Undo Mark Cursor, and clear all Marks. Choosing
Mark
Cursor
will create a note showing the cursor value at the cursor position.
Calibration
:
Calibration
:
Selecting Calibration is for calibrating a
sensor (adjusting the sensitivity).
When you require sensor calibration prior to
measurement, this procedure is used.
For example, an acoustic calibrator usually
generates a 94dB calibration signal at 1 kHz,
therefore you must enter the signal
frequency by 1000 Hz, then select “Acquire
Signal” to get the reading in the “old value”
Содержание Fieldpaq