Alarm texts for the digital inputs
To all digital inputs a neutral or specific alarm message can be assigned. A neutral alarm message in-
dicates the alarm, the channel and the address of the device triggering the alarm. In comparison, a
specific alarm message (medical) signals a fixed pre-programmed alarm, e.g. "oxygen". Assign the in-
puts according to the table "Neutral and specific alarm messages"
The TMK-SET PC software can be used to assign other messages to individual or all digital inputs.
The alarm messages in the following table are sent to other TM… or MK… panels via the BMS bus
and displayed there in plain text format. If freely programmable alarm messages need to be dis-
played on a different TM… or MK…, the same alarm messages must have been programmed in the
displaying device.
Specific alarm messages
These messages contain details about medical gases and BSV systems. Alarm messages for medical
gases are signalled by the red "ALARM" LED and the buzzer sound. The buzzer sound can be set to
mute (acknowledged). The buzzer sounds again every 12 minutes as a reminder. Individual settings
can only be made using the TMK-SET software.
Neutral and specific alarm messages
Inputs
Neutral alarm messages
"FuncNeutral"
Specific alarm messages
"Function: Medical"
LED
IN1
Alarm: Address/channel XXX/01
Alarm: Oxygen
Alarm (red)
IN2
Alarm: Address/channel XXX/02
Alarm: Vacuum
IN3
Alarm: Address/channel XXX/03
Alarm: Nitrous oxide
IN4
Alarm: Address/channel XXX/04
Alarm: Compressed air 5 bar
IN5
Alarm: Address/channel XXX/05
Alarm: Compressed air 8 bar
IN6 Alarm:
Address/channel XXX/06
Alarm: Nitrogen
IN7
Alarm: Address/channel XXX/07
Alarm: CO2
IN8
Alarm: Address/channel XXX/08
Alarm: BSV battery operation
Warning (yel-
low)
IN9 Alarm:
Address/channel
XXX/09 Alarm:
BSV
overload
IN10
Alarm: Address/channel XXX/10
Alarm: BSV converter failure
IN11
Alarm: Address/channel XXX/11
Alarm: BSV fault
IN12
Alarm: Address/channel XXX/12
Alarm: BSV test run
IN13
Alarm: Address/channel XXX/13
Alarm: BSV mains operation
IN14
Alarm: Address/channel XXX/14
Alarm: Failure air conditioning
IN15
Alarm: Address/channel XXX/15
Alarm: OP light battery operation
IN16
Alarm: Address/channel XXX/16
Alarm: Sat OP light battery opera-
tion