Operation
51
TM800_D00165_00_M_XXEN/06.2015
7.2.2
Display during fault condition
A warning or an alarm message exists.
Depending on the type of fault, either the yellow LED "Warning" or the red LED "Alarm" will
light up. The green LED "Normal" no longer lights up.
The buzzer sounds simultaneously. If the cause of fault cannot be remedied immediately, the
buzzer can be muted by pressing the "
" (6) button.
The LC display shows details about the message.
– Line 1: Standard display. "System": Enter user-defined text here: "Intensive care unit 03"
– Line 2…3: Message text, measured value (if available)
– Line 4
Status line
xx
= Sequence number of the message displayed
yy
= Number of pending messages
= Message text page, in this case page 1
zzz
= Insulation fault location or test currently in progress (see table).09:50=
Time (example)
During the insulation fault location or test process, the following displays are possible:
Only when the external bus is in "Off" position:
zzz Meaning
EDSa
EDS insulation fault location in progress (automatic)
EDSp
Continuous EDS insulation fault location in progress
EDSs
Single-pass EDS fault location in progress
EDS
EDS insulation fault location process has been completed, the current measuring
sequence is still running.
TEST
Test is running. The message "TEST" flashes if the message being displayed has been
caused by the test.
noMA
No master on the internal bus.
MAST
Device is "substitute master" on the internal bus.
Intensive care unit 03
Insulation fault
Measured value 43 k
Ω
xx/yy zzz 09:50