ISOMETER® Modbus register assignment
39
isoHV525_D00297_04_M_XXEN/05.2018
The alarm type is coded by the bits 0 to 2. Bits 3, 4 and 5 are reserved and always have
the value 0. Bit 6 or 7 is set when an internal or external test has been completed. Other
values are reserved. The complete byte is calculated from the sum of the alarm type and
the test type.
8.1.2.3
R&U = Range and unit
X
X
X
X
X
1
1
0
Reserved
X
X
X
X
X
…
…
…
Reserved
X
X
X
X
X
1
1
1
Reserved
Te
st
0
0
X
X
X
X
X
X
No test
0
1
X
X
X
X
X
X
Internal test
1
0
X
X
X
X
X
X
External test
Bit
7
6
5
4
3
2
1
0
Meaning
Unit
-
-
-
0
0
0
0
0
Invalid (init)
-
-
-
0
0
0
0
1
No unit
-
-
-
0
0
0
1
0
Ω
-
-
-
0
0
0
1
1
A
-
-
-
0
0
1
0
0
V
-
-
-
0
0
1
0
1
%
-
-
-
0
0
1
1
0
Hz
-
-
-
0
0
1
1
1
Baud
-
-
-
0
1
0
0
0
F
-
-
-
0
1
0
0
1
H
-
-
-
0
1
0
1
0
°C
-
-
-
0
1
0
1
1
°F
-
-
-
0
1
1
0
0
Second
-
-
-
0
1
1
0
1
Minute
-
-
-
0
1
1
1
0
Hour
-
-
-
0
1
1
1
1
Day
-
-
-
1
0
0
0
0
Month
Ra
n
g
e
o
f v
a
li
d
ity
0
0
X
X
X
X
X
X
Actual value
0
1
X
X
X
X
X
X
The actual value is lower
1
0
X
X
X
X
X
X
The actual value is higher
1
1
X
X
X
X
X
X
Invalid value