26
4.4 Connecting NO Gas Cylinders
1.
Using the on-screen Quick
Start Guide (QSG)
instructions set up the
system as follows:
NOTE: for experienced users, the
“Skip to System Tests” button will
bypass the on-screen QSG and
skip to system setup tests).
Alternatively the “skip to dose”
button will bypass both on-screen
QSG and the system test.
See Chapter 6.3 for detailed
instructions for using the touch
screen interface.
2.
Check that the
NOxBOX
i
NO cylinder
concentration matches
the value shown on the
system screen. This is
the value upon which
delivery calculations
will be based.
Do not use
any supply cylinder that
does not match this
value
.
If there is a problem,
contact your Service
Engineer to resolve before
attempting to deliver NO to
a patient.
3.
Check the regulator
sealing surfaces and
connector thread. If your
model has non-metal
sealing elements, ensure
that they are present and
undamaged, showing no
cracks, pock-marks, etc.