3.5.3 Trigger Sensitivity
When using synchronised ventilator modes, the net change of gas volume flow caused by the
NOxBOX
i
may trip the trigger flow setting. Always check the trigger sensitivity of the ventilator after
connecting the NOxBOXi to the patient circuit.
3.5.4 Maximum NO Dose Delivery
The NOXBOX is capped at maximum dose capability of 80 ppm in software.
Additionally, the maximum flow capability of the NOxBOX
i
is physically limited. This will cause
some dose limitations at higher ventilator flow rates.
The NOxBOX
i
is compatible with a wide range of NO supply concentrations, note that the physical
flow limitation has less effect on the 1000 ppm cylinder supply compared to the 400 ppm supply.
See Figure 3.5.4 1: Graph of NOxBOXi Maximum Dosing Capability.
NOxBOX
i
Max. Dosing Capability
See Figure 3.5.4-1: Graph of NOxBOX
i
Maximum Dosing Capability
NO Dose (
ppm)
Constant Flow Minute Volume (L/min)
0
10
20
30
40
50
60
70
80
0
10
20
30
40
50
1000 ppm
900 ppm
800 ppm
500 ppm
450 ppm
400 ppm
300 ppm
250 ppm
225 ppm
200 ppm