4.3 Setup Instructions
To setup the
NOxBOX
i
system and prepare it for use:
1.
Ensure that the system
trolley brakes are engaged
during setup and in use.
2.
Insert the monitor end of
the power cord into
the power socket at the
rear of the monitor, and push
it in until the locking
mechanism clicks.
3.
Connect the power supply
plug to a hospital-grade
electrical outlet that is
backed up by an
emergency power generator.
4.
Secure each cylinder to
the trolley using the straps
provided. The maximum
NO cylinder that the trolley can
accommodate is 25 litres
water capacity (max. height:
1.2 m). Additional space
has been provided on
the trolley for a small oxygen
cylinder if desired.
5.
Select the number of
available NOxBOX
i
NO
supply cylinders that are
being setup on the system
(one or two).
6.
Ensure that the system is set
to intelligent delivery; the
mode selection valve should
be in the vertical position, as
shown here.
7.
The manual control valve
immediately below this switch
should be set appropriately
for the patient (see Section
6.12 for details).