Follow these precautions when cleaning the NOxBOX
i
System:
•
The sensors inside the system can be damaged by exposure to alcohol or volatile organic
compound (VOC) spray or fumes. Do not use alcohol- or VOC- based cleaners on this
system. Alcohol-free instrument cleaning wipes are available from Bedfont that are
suitable for cleaning the NOxBOX
i
system monitor.
•
Clean external surfaces and the touchscreen panel using a damp, soft, lint-free cloth.
•
Do not spray or splash cleaning fluids directly onto the system surfaces. Apply cleaning
fluids with a clean soft lint-free cloth, and avoid excessive wetting of the system. Take
extra care around all connectors to prevent fluid from entering and damaging the system.
•
Only use mild soap and water or an approved disinfectant for cleaning the NOxBOX
i
System. Always follow manufacturer’s recommendations when using any disinfectant.
•
Do not clean the system with organic solvents, glass cleaner, petroleum-based solvents,
alcohol, or acetone-based solvents.
•
Do not autoclave or gas-sterilize the system.
•
Do not use any abrasive cleaning pads or cleansing fluids on the NOxBOX
i
system, and
take particular care not to scratch the touchscreen panel.
•
To prevent damage to the touchscreen, avoid excessive pressure when cleaning the touch
screen panel.
•
Ensure that the system is completely dry before returning it to use after cleaning.
Alcohol-free instrument wipes approved for use with the NOxBOX
i
system are available
from Bedfont Scientific Ltd.
7.4 Calibration
CAUTION: Temperature may affect the accuracy of the NOxBOX
i
monitoring function. The instrument should be calibrated at the
temperature at which it is expected to be used.
CAUTION: In order to correctly calibrate the NOxBOX
i
, ensure that
the pump flow is checked and if necessary adjusted prior to
performing sensor calibrations.
CAUTION: Always perform a sensor zero calibration before
performing the full-scale calibration.
7.3 Cleaning the NOxBOX
i
System
WARNING: Ensure that the unit is turned off and the power cord is
completely disconnected before cleaning the
NOxBOX
i
System.