MPS 500 User Manual Rev.2.1
BeamTec GmbH
Wolfgang-Paul-Str. 4, 89081 Ulm, Germany
http:
, e-mail: [email protected] tel.: +49 731 146620-0
47
NOTICE: The minimum brightness value is limited, in order to prevent the total
extinction of the display..
NOTE: It is recommended to set the backlight to minimum, where you still can
comfortably read the parameters., otherwise there will be a corresponding
reduction of display lifetime.
Use arrow button
▼
to navigate to the Contrast settings
– Fig. 4.4.4
Fig.4.4.4
Use the rotary encoder to adjust the brightness. The range can be set from 35 ....
100%.
NOTICE: Like the backlight, the minimum value is limited in order to avoid complete
shutdown of the display.
4.4.2
INTERLOCK
TYPE
OPTION
MPS500 is equipped with external interlock – see section: 3.3.5
However there is also an internal interlock which can be used to operate the unit. This
interlock is triggered via the vacuum measurement.
NOTICE: The interlock can be activated/deactivated only from within this menu. To
set the triggering value, please refer to the Vacuum gauge settings – see section: 4.4.4
Use arrow buttons
▲
or
▼
to navigate and point to the Interlock Type function. Press
MENU/ENTER button to proceed and continue with arrow buttons
▲
or
▼ to
activate/deactivate the internal interlock – Fig. 4.4.4
Fig.4.4.4
Brightness: 80%
Contrast: 65%
<
Interlock Type:
EXT + INTERNAL
<