MPS 500 User Manual Rev.2.1
BeamTec GmbH
Wolfgang-Paul-Str. 4, 89081 Ulm, Germany
http:
, e-mail: [email protected] tel.: +49 731 146620-0
45
-
Gateway,
-
Port,
-
DHCP,
-
TCP Server Port
-
Status
NOTE: Status option has to be set to “ON” before obtaining DHCP address
2.
RS232/485
-
Baudrate ( 2400, 4800, 9600, 19200, 38400, 57600, 115200 ),
NOTE: the other parameters of this interface are fixed, and take the following
values: 8 data bits, 1 stop bit, No parity
3.
Bluetooth
-
Baudrate ( 2400, 4800, 9600, 19200, 38400, 57600, 115200 ),
NOTE: the other parameters of this interface are fixed, and take the following
values: 8 data bits, 1 stop bit, No parity
IMPORTANT: Because of EMC issues, MPS500 chassis is strongly shielded.
This brings effects on the limited signal strength of the internal Bluetooth
module and limits working distance to 5m only.
4.4.0.3
UNIT
ADDRESS
O
PTION
Unit Address is the individual address which has to be assigned to the device and is
used for serial communication protocol – please refer to the separate manual. It does
not matter which communication interface has been chosen.
Unit can take an address from 0…255.
Use arrow buttons
▲
or
▼
to navigate and point to the Communication function.
Press MENU/ENTER button to proceed and continue with arrow buttons
▲
or
▼ to
configure the particular interface – Fig. 4.4.1