Mains power.
Improperly grounded devices can be extremely dangerous in the event of a
fault. Use three-wire mains or extension cables with protective ground only.
Plug the mains cable into wall sockets with protective ground only.
1. Connect the European appliance connector of the mains cord with the mains
connection of the device
2. Connect the plug of the mains cable with the wall socket
Note:
If the device is installed in a switching cabinet, the mains power can be supplied via a
switchable central power distributor.
3.3.3.
G
ROUNDING
Grounding screw (Fig. 3.3.0, the reference H) should be used to connect the MPS500
with the main grounding system in which it operates. It is recommended to use a
cable with a minimum section of 2.5 mm2
If required, connect the vacuum system ground from the earthing screw using the
protective conductor.
3.3.4.
H
IGH
V
OLTAGE
C
ONNECTORS
HV1
AND
HV2
The power supply is equipped with two HV outputs of N-50/2-0/G1.07
type.
Output
configuration is set by the software user interface. Please refer to chapter 4.3.0
1
DANGER
!