General - Warnings 61
Communication may also be started from the following main functions in MC6-
Ex:
Calibrator
,
Documenting Calibrator
and
Data Logger
.
Depending on active/inactive options, some communication protocols may be
disabled (grayed).
MC6-Ex's
Communicator
is primarily meant for viewing and editing the configu-
ration of Digital Communication Instruments. However, when invoking communi-
cation from
Calibrator
,
Documenting Calibrator
or
Data Logger
, you most
likely plan to
select a Variable or Parameter
(e.g. Primary Value) for calibration
or data logging instead of doing full configuration for the instrument. MC6-Ex, of
course, supports this functionality too.
See also
HART Device Description Specifics
on page 69.
Selecting a communication protocol in Calibrator,
Documenting Calibrator and Data Logger.
W
ARNINGS
Configuring/calibrating an instrument that is part of a live segment is possible in HART
and FOUNDATION Fieldbus only. When working in a live segment, first make sure, the
control loop the instrument is part of is set to manual. Follow the guidelines available in
the instrument's own manual.
Beamex cannot be held responsible for any damages caused by connecting MC6-Ex to a
live factory fieldbus segment.
Using MC6-Ex to change the parameters of an instrument
may result in discrepancies: A fieldbus host control system
may mirror all instrument parameters in its permanent da-
tabase. In such a case, when returning an instrument with
changed parameters to a live segment, ensure that the pa-
rameters are also available in the control system's perma-
nent database. Also verify that the new parameters do not
result in an unstable control loop.
Содержание MC6-Ex
Страница 2: ......
Страница 37: ...Things discussed in this part Presenting the Meter and how to take it into use Part 3 METER ...
Страница 60: ...52 MC6 Ex User Manual Part 5 Documenting Calibrator ...
Страница 96: ...88 MC6 Ex User Manual Part 9 Additional Information ...
Страница 112: ...104 MC6 Ex User Manual Appendix ...