Additional Information - User Defined Steps / Calibration Points 85
U
SER
D
EFINED
S
TEPS
/
C
ALIBRATION
P
OINTS
These are available for instruments in
Documenting Calibrator
and also in
Cal-
ibrator
's
Step
tool. Use either
Calibration Points
or
Step Definition
button to
modify its setting. In the opened window, browse to
User Test Points
page.
The configuration consists of two (or more) pages as shown in the pictures to the
right of this text. The first page is for general definitions and the second for enter-
ing steps / calibration points. If you enter more points than can be shown on one
page, a further page is automatically added.
There are no "rules" in what kind of values are entered. They may be outside
given point data entry range and need not to be in any strict order. See the lower
picture to the right.
Also:
The row number is a button. It opens a window where you may delete the
current point or optionally add a row either before or after the current
point.
An added point gets a default value: Average value of preceding and fol-
lowing point.
1
st
configuration page.
2
nd
configuration page.