28
6-847718-00 Rev. B00
Holes for entrance of
low voltage field wiring
Two choices of
conduit holes for
entrance of AC
mains power
top entry
side
entry
For top entry, remove
shield for access.
Reinstall shield after
connection is made
Figure 22: Wire Routing for Area Alarms
Figure 21: Wire Routing for Master / Combo Alarms
Two holes or knockouts are provided for
entrance of AC mains power. One hole
on left side and one hole on top left of
alarm panel back box.
NO OTHER HOLES SHOULD BE
PUNCHED OR USED.
Several additional holes or knockouts are
provided on right side of top panel and
bottom for entrance of low voltage field
wiring(Figures 21 and 22).
Two choices of
conduit knockouts
for entrance of AC
mains power
top
entry
side
entry
Knockouts for entrance of
low voltage field wiring
Knockouts for entrance of
low voltage field wiring
Wire Routing
Pull Wire (Cont.)
Inst
allation
Procedures
Содержание TotalAlert 2
Страница 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Страница 2: ......
Страница 22: ...20 6 847718 00 Rev B00...
Страница 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Страница 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Страница 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Страница 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Страница 130: ...128 6 847718 00 Rev B00...
Страница 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Страница 143: ......
Страница 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...