
41
6-847718-00 Rev. B00
1. Remove sensor module from shipping
carton.
2. Insert sensor into valve opening in
pressure/vacuum service rough-in
(Figure 33). An indexing pin provides
for proper orientation and prevents
gas service cross-connection. Push
sensor all the way in and secure with
(2) #6-32 x 1-1/4” screws. Repeat this
process for all sensors within alarm
panel.
3. The digital display module(s) located
on front panel assembly are factory
wired with a harness and polarized
two-pin connector for each sensor
module. Each digital display module
is labeled with the gas type. Match
gas type of each sensor to gas type of
each digital display module and join
polarized connectors of sensor and
digital display module harnesses.
NOTE:
Remove blue protective netting from sensor
connector.
Local Sensor
Figure 33: Local Sensor Mounting
Finish
Inst
allation
Procedures
Indexing Pin
Indexing Hole
Screws
NOTE:
Pressure/vacuum service rough-ins have been
installed in a specific order to optimize sensor
module wire routing.
Содержание TotalAlert 2
Страница 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Страница 2: ......
Страница 22: ...20 6 847718 00 Rev B00...
Страница 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Страница 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Страница 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Страница 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Страница 130: ...128 6 847718 00 Rev B00...
Страница 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Страница 143: ......
Страница 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...