42
6-847718-00 Rev. B00
Inst
allation
Procedures
1. Remove sensor module from shipping
carton. Each sensor module is
shipped with a wiring pigtail that has a
two-pin polarized connector.
2. Wire nut pigtail to field installed wiring
inside sensor mounting box. Note
polarity of wiring and corresponding
field wiring color or number for later.
3. Join polarized connector in sensor
mounting box to mating connector on
sensor module (Figure 34).
4. Insert sensor into valve opening in
sensor rough-in. An indexing pin
provides for proper orientation and
prevents gas service cross-
connection. Push sensor all the way
in and secure with (2) #6-32 x 1-1/4”
screws.
5. Tuck excess wiring into open space
behind sensor.
NOTE:
Remove blue protective netting from sensor
connector.
Figure 34: Remote Sensor Module Connection
Remote Sensor
Valve opening
Sensor
Polarized
connector
Finish
NOTE:
Do not ground shield drain wire at sensor.
Содержание TotalAlert 2
Страница 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Страница 2: ......
Страница 22: ...20 6 847718 00 Rev B00...
Страница 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Страница 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Страница 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Страница 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Страница 130: ...128 6 847718 00 Rev B00...
Страница 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Страница 143: ......
Страница 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...