49
6-847718-00 Rev. B00
Wiring Schematic 4: Relay Board
W
iring
Schematics
Breakout
board B
Breakout
board A
Relay
board A
Relay
board B
Dry normally-closed
output relay contacts
NOTE:
Output relay contacts are dry normal-
ly closed type. Contacts open upon
designated alarm activation.
Pressing MUTE
on Annunciator
Module DOES NOT reset relay.
Relay deactivates when alarm condi-
tion is corrected.
Relay contact ratings are 2 A @ 30
VDC/0.5 A @ 125 VAC.
NOTE:
Field wiring cable shields must be grounded at
only one end, inside alarm panel back box. Refer
to page 39 for details.
SIGNAL A1
SIGNAL A2
SIGNAL A32
SIGNAL B1
SIGNAL B32