background image

 

 

 

 

MAK 3003 EASI Ultrasampler Controller Type 6970-31, Configuration, Software version EASI 1.00, SAK 200311 

All rights reserved. Subject to change without prior notice. 

 

No part of this document may be reproduced, processed or distributed   
in any form or by any means without the prior written permission of   
BARTEC BENKE

 

Copyright © 2020 by  

BARTEC BENKE  

Schulstraße 30 

D-94239 Gotteszell 

 
 

Содержание MAK 3003

Страница 1: ...4239 Gotteszell Deutschland Phone 49 0 9929 301 0 Telefax 49 0 9929 301 112 E Mail gotteszell bartec benke de Internet www bartec benke de MAK 3003 EASI Ultrasampler Controller Type 6970 31 Configurat...

Страница 2: ......

Страница 3: ...00 SAK 200311 All rights reserved Subject to change without prior notice No part of this document may be reproduced processed or distributed in any form or by any means without the prior written perm...

Страница 4: ......

Страница 5: ...______________________ 9 4 Menu____________________________________________ 10 4 1 System Info _______________________________________ 10 4 2 Diagnostic ________________________________________ 11 4 3...

Страница 6: ...BENKE accepts no liability for any damage resulting from non ob servance of the safety regulations or from non compliance with the operating instructions or operating conditions Secondary damage is ex...

Страница 7: ...er slightly from the actual design of the device BARTEC GmbH reserves the right to make technical changes at any time BARTEC GmbH is under no circumstances responsible or liable for any indi rect or c...

Страница 8: ...ians Make sure that the information and operating conditions listed by the man ufacturer are followed Follow the instructions on operating and servicing the devices If you notice any damage or deterio...

Страница 9: ...cessed measurement values of the flow level meter Switching on the system and operating the vehicle system depend on the spe cific vehicle type and the operating instructions applicable thereto 3 1 Ul...

Страница 10: ...menus and also to enter numerical values To accept the numerical values entered and to exit the input dialog use the two buttons on the right side of the display By pressing the green button a menu ca...

Страница 11: ...ning the EASI Wheel left and right The selected submenu is marked by a colored bar You can then open the selected menu by pressing the EASI Wheel or pressing the green button If the menu contains subm...

Страница 12: ...d list that you want to accept The selected parameter is marked by a bright bar 2 The currently selected parameter is accepted by pressing the EASI Wheel or pressing the green button Numerical entries...

Страница 13: ...Changing the configuration is protected by entering a password After confirming the selected field enter password the request for the pass word appear As with numerical entries 3 2 1 2 the password is...

Страница 14: ...contains submenus for configuring the system and calling up various functions 4 1 System Info This menu shows the A Nr the version and the selected hardware type System Info Display Typ type name e g...

Страница 15: ...voltage and power inputs for FLM are displayed Diagnostics Run Motor 1 The sampling motor 1 will start for a few seconds Run Motor 2 The sampling motor 2 will start for a few seconds Inputs Outputs By...

Страница 16: ...s customised to the respective operating conditions and the installed hardware by entering various parameters Changing the configuration is protected by entering a password 3 2 2 In the following para...

Страница 17: ...Display Display Contrast In this menu the contrast of the display can be changed dark middle light Standby time Input the Standby time Input in sec Time Date Enter the date and the current time The s...

Страница 18: ...tected against the ve hicle power supply voltage to ensure to all control tasks Therefore connec tions to other potentials e g chassis ground reference or reverse voltages in the plus branch are alway...

Страница 19: ...ilk is relayed to the ULTRASAMPLER small medium quantity With preset quantities below this value the small pumping rate is relayed to the ULTRASAMPLER with other quantities the me dium rate is relayed...

Страница 20: ...value if sampling vial filling volume deviates from the set value Pre Flushing From Estimate In case of intake quantities set quantities lower than the value en tered here no pre flushing occurs L Flu...

Страница 21: ...he time period for temperature monitoring Valid until Time until when the milk is monitored for compliance with the maxi mum allowed temperature Min allowed temp Lower temperature limit for milk intak...

Страница 22: ...311 No Pumping is not interrupted if the maximum allowed temper ature is exceeded Only a warning message is shown 4 3 9 Cleaning Cleaning Cleaning Id The cleaning Id will be recorded Min Hot Wash Temp...

Страница 23: ...All current errors are displayed here Error history All current errors since the last restart are displayed here Status All current statuses are displayed here Status history All current statuses sin...

Страница 24: ...Display Contrast dark middle light Standby time Time Date Time Date Demomode Demomode off Demomode on Outputs Output 1 Output 2 Inputs Input 1 Resting state high negative switching low positive switch...

Страница 25: ...ng Steps Minimum Pulse Steps Minimum Pulse Frequency Motor 2 Operation Mode Not installed Single sample Tour sample Vial Volume Overfill Tolerance Start Delay Calibration Factor Pre Flushing From Esti...

Страница 26: ...MAK 3003 EASI Ultrasampler Controller Type 6970 31 Configuration Software version EASI 1 00 SAK 200311...

Отзывы: