MAK 3003 EASI Ultrasampler Controller Type 6970-31, Configuration, Software version EASI 1.00, SAK 200311
Calibration Factor
The pumping hose calibration factor indicates how many µl are filled
per sampler rotation. (Enter 270 µl. Change the value if sampling vial
filling volume deviates from the set value.)
Pre Flushing From Estimate
In case of intake quantities (set quantities) lower than the value en-
tered here, no pre-flushing occurs. [L]
Flushing Steps
Number of flushing steps when milk is detected in the pipe.
Minimum Pulse Steps
Minimum number of motor steps.
(10)
Minimum Pulse Frequency
Minimum rpm of the sampler. If this rpm value is not reached, this
triggers intermittent mode.
(50)
4.3.8
Milk temp. limits
The milk to be collected can be monitored for compliance with a max. allowed
temperature depending on time of day. To this end, you can enter three times
with the corresponding temperature limits.
If the temperature limits are exceeded, a warning message is displayed and
the pump is stopped, if this is configured (see below).
Milk temp. limits
Delay Temp watch
Delay of the temperature measurement after the system detects me-
dium. [s]
Temp. warning limit
Temperature threshold. The value determines how many °C before
reaching a temperature limit, a warning is displayed. [°C]
Time period 1
Activate or deactivate the time period for temperature monitoring
Valid until
Time until when the milk is monitored for compliance with the maxi-
mum allowed temperature.
Min. allowed temp.
Lower temperature limit for milk intake during this time period [°C]
Max. allowed temp.
Upper temperature limit for milk intake during this time period [°C]
Interrupt Pumping
Yes:
Pumping is interrupted if the maximum allowed temperature
is exceeded.
No:
Pumping is not interrupted if the maximum allowed temper-
ature is exceeded. Only a warning message is shown.
Time period 2
Activate or deactivate the time period for temperature monitoring
Valid until
Time until when the milk is monitored for compliance with the maxi-
mum allowed temperature.
Min. allowed temp.
Lower temperature limit for milk intake during this time period [°C]
Max. allowed temp.
Upper temperature limit for milk intake during this time period [°C]
Interrupt Pumping
Yes:
Pumping is interrupted if the maximum allowed temperature
is exceeded.
No:
Pumping is not interrupted if the maximum allowed temper-
ature is exceeded. Only a warning message is shown.
Time period 3
Activate or deactivate the time period for temperature monitoring
Valid until
Time until when the milk is monitored for compliance with the maxi-
mum allowed temperature.
Min. allowed temp.
Lower temperature limit for milk intake during this time period [°C]
Max. allowed temp.
Upper temperature limit for milk intake during this time period [°C]
Interrupt Pumping
Yes:
Pumping is interrupted if the maximum allowed temperature
is exceeded.